Overpressure Limit; Density; Viscosity; Pressure Tightness - Endress+Hauser Liquiphant FTL63 Operating Instructions Manual

Point level switch for liquids specifically for the food and life sciences industries
Hide thumbs Also See for Liquiphant FTL63:
Table of Contents

Advertisement

Technical data
58
14.4.4
Process pressure range of the sensors
PN
[psi]
[bar]
1450
100
928
64
32
0
-50
-58
1
PN: 64 bar (928 psi) for a maximum of 150 °C (302 °F), refer to "Process connections" section for exceptions
14.4.5

Overpressure limit

• PN = 64 bar (928 psi): overpressure limit = 1.5 · PN maximum 100 bar (1 450 psi)
depending on the selected process connection
• Membrane burst pressure at 200 bar (2 900 psi)
The device function is limited during the pressure test.
Mechanical integrity is guaranteed up to 1.5 times the process nominal pressure PN.
14.4.6

Density

Liquids with density > 0.7 g/cm³ (43.7 lb/ft³)
Switch position > 0.7 g/cm³ (43.7 lb/ft³), order configuration
Liquids with density 0.5 g/cm³ (31.2 lb/ft³)
Switch position > 0.5 g/cm³ (31.2 lb/ft³), can be configured via DIP switch
Liquids with density > 0.4 g/cm³ (25.0 lb/ft³)
• Optionally available to order
• SIL for defined media and process parameters on request
• Fixed value that cannot be edited
The function of the DIP switch is interrupted
For information on medium differentiation/density detection: Documentation
Liquiphant density (FEL60D) with density computer FML621 (Endress+Hauser
website
www.endress.com
14.4.7

Viscosity

≤ 10 000 mPa⋅s
14.4.8

Pressure tightness

Up to vacuum
In vacuum evaporation plants, select the 0.4 g/cm³ (25.0 lb/ft³)/ density setting.
14.4.9

Solids contents

⌀ ≤ 5 mm (0.2 in)
1
0
32
→ Downloads)
Liquiphant FTL63
T
+150
[°C]
p
[°F]
+302
A0051481
Endress+Hauser

Advertisement

Table of Contents
loading

Table of Contents