8.3 Preventive Maintenance and Servicing

Service Life
This device has an anticipated minimum service life of 12 years, provided
that the service is performed according to the manual.
Note:
A failure to undertake routine preventive maintenance and product
service in line with recommendations in the product manual, or misuse and/or
abuse of the device, may reduce the operating life expectancy.
To ensure proper operation of the eAdvantage®, regular inspection and checking
of the device and accessories for correct function should be undertaken by a
responsible member of sta . It is recommended that a preventive maintenance
record be maintained for each unit.
It is recommended that routine preventive maintenance be carried out, and
the eAdvantage® mixer be returned to O-Two™ Medical Technologies (or its
authorized service center) for maintenance and service every 2 years as follows:
Monthly checking
This check is to ensure that all of the accessories and device components are
present, the nitrous oxide and oxygen cylinders are full and that the device is
in working order by turning ON the gas supplies and the device and confi rming
that the self-diagnostic does not show any issues.
Level II service
The device shall be returned to the manufacturer or a service center authorized
37