O-Two eAdvantage User Manual page 42

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Table of Contents

Advertisement

Electromagnetic Immunity
Immunity test
Electrostatic discharge
(ESD) IEC61000-4-2
Electrical fast transient/
Burst IEC61000-4-4
Surge IEC61000-4-5
Voltage Dips
IEC61000-4-11
Voltage Interruptions
IEC61000-4-11
Power frequency magnetic
fi eld IEC61000-4-8
Conducted RF
IEC61000-4-6
Radiated RF EM fi elds
IEC61000-4-3
Proximity fi elds from RF
wireless communications
equipment IEC61000-4-3
Use of accessories, transducers, and cables other than those specifi ed or provided by
the manufacturer of eAdvantage® could result in increased electromagnetic emission or
decreased electromagnetic immunity of eAdvantage® and result in improper operation.
The eAdvantage® is intended for use in the electromagnetic environment
in which radiated RF disturbances are controlled. The user can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (including peripherals such
as antenna cables and external antennas) and the eAdvantage®.
Portable RF communication equipment should be used no closer than
42
WARNING
Use of eAdvantage® adjacent to or stacked with other
equipment should be avoided because it could result
in improper operation of the device.
WARNING
WARNING
30 cm (12 inches) to any part of the eAdvantage®.
Immunity test level
± 8 kV contact
± 2, ±4, ±8, ±15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
±0.5, ±1 kV line to line;
±0.5, ±1 kV, ± 2 kV line to earth
0.5 cycle, 1 cycle,
25 (50Hz)/30 (60Hz) cycles
250 cycle (50Hz),
300 cycle (60Hz)
30 A/m (50/60 Hz)
6Vrms: 150kHz to 80MHz in ISM bands
6Vrms: 150kHz to 80MHz in ISM bands
3 V/m @ 80 MHz to 2.7 GHz
IEC 60601-1-2 Chapter 8.10 Table 9
eAdvantage User Manual