O-Two eAdvantage User Manual page 32

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Table of Contents

Advertisement

Alarm/Warning priorities with corresponding visual/ audible alarms
Symbol
Name
Sensor
error (U13
and/or U14
sensors
failure)
No O
gas
2
O
input
2
≤ 20 PSI
Low O
input
pressure
(45-21 PSI)
High O
input
pressure
≥ 80 PSI
No N
O
2
gas input
≤ 20 PSI
Low N
O
2
input
pressure
(45-21 PSI)
High N
2
input
pressure
≥ 80 PSI
Empty
Battery
Low
Battery
32
Cell
Priority
#
A
High
B
High
2
B
Medium
2
B
High
C
High
C
Medium
O
C
High
High
Low
On Screen
Alarm delay
Visual Alarm
Solid symbol
Directly
with fl ashing red
warning symbol
Solid symbol
Directly
with fl ashing red
warning symbol
Solid symbol fl ashing
Directly
down with yellow
warning symbol
Solid symbol
Directly
with fl ashing red
warning symbol
Solid symbol
Directly
with fl ashing red
warning symbol
Solid symbol fl ashing
Directly
down with yellow
warning symbol
Solid symbol
Directly
with fl ashing red
warning symbol
Solid symbol with a
Directly
fl ashing red warning
The solid yellow
N/A
symbol with a fi xed
yellow warning symbol
Audible Alarm
1 Burst with 10
pulses each, repeat
every 7.5 seconds
1 Burst with 10
pulses each, repeat
every 7.5 seconds
1 Burst with 3 pulses
each, repeat every
20 seconds
1 Burst with 10 pulses
each, repeat every 7.5
1 Burst with 10
pulses each, repeat
every 7.5 seconds
1 Burst with 3
pulses each, repeat
every 20 seconds
1 Burst with 10
pulses each, repeat
every 7.5 seconds
1 Burst with 10
pulses each, repeat
symbol
every 7.5 seconds
eAdvantage User Manual
seconds
N/A

Advertisement

Table of Contents
loading

This manual is also suitable for:

01eq3000

Table of Contents