Software Update - O-Two eAdvantage User Manual

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Table of Contents

Advertisement

After confi rming the passcode, select the software update option.
Insert a USB key with the new software revision to update, or select the 'Cancel'
symbol to exit without updating.
Upon inserting the boot loader USB key into the eAdvantage® system, the
confi rmation indicator light (J) will start fl ashing, indicating the need to confi rm
the selection manually before initiating the software update process.
To confi rm and start the update, press the Control Knob (K). The boot loader
will then automatically proceed with the software update. During this process,
the confi rmation indicator light (J) will remain ON, signaling that the new
software is uploading into the device.
Note: If both the software revisions in the device and the USB Key are the
same, the device will cancel this process and restart the mixer without writing
software from the USB key into the device.
Once the update is completed, the device will shut down and restart
automatically. Once it's on, the technician must confi rm that the software
version shown on the startup page (Fig. 1) is identical to the revision number
written on the new USB key.