5.4.4. Stored Events

Note: The "Data display" is not protected by a passcode. To display the stored
events, follow the setting mode steps below and press the "Data display" cell
and the screen will change to:
The mixer will store any event including all alarms and their related actions,
turning the device ON or OFF, pressing the Oxygen Flush key, setting changes
after confi rmation, activating and canceling the alarm silence, calibration results,
software updates as well as locking and unlocking the screen.
All stored events are saved, per event, up to a maximum of 2048 events after
which the software will start overwriting the older stored data starting from the
oldest event. The contents of the stored events can be maintained for about 7
weeks after a total loss of both power supply mains and internal battery.
To view the data, users must turn the control Knob (K) clockwise to move the
stored data down and display older events or counterclockwise to display
previous events. Scrolling the Control Knob will move the events up or down by
one event per turn.
If multiple events occur at the same time, their events will be saved sequentially
with no particular priority.
Once completed the user can exit the mode by pressing the Cancel Key.
Note: Individual stored events cannot be deleted by the user.