7.2 Flow Valve Leak Alarm

The eAdvantage® system includes a fl ow valve leak detection feature that
functions when the unit is supplied with gas.
While the device is ON, it continually monitors the outputs from both fl ow
sensors. This monitoring occurs during the standby mode or for 10 seconds
following the last patient-demanded breath.
If either of the fl ow sensors detects any unexpected fl ow (suggestive of a leak),
the device will promptly display leak alarms on the screen, alerting the user to
the issue.
Note: It's important to note that the leak detection feature will be inactive when
both input gases are turned OFF (0 PSI). This means the detection feature only
operates when there is gas pressure present in the system.
eAdvantage User Manual
Cell
Priority
#
% @
Medium
D
or high below
27% of O
% @
2
D
Medium
D
High
O
2
Alarm
delay
After 3
consecutive
%
breaths
2
Solid symbol fl ashing
After 3
consecutive
breaths
10 seconds
after turning
A solid symbol with a
the device
fl ashing red Warning
ON or the last
demanded
breath
On-Screen
Visual Alarm
Follow the
visual alarm logic
depending on the
alarm priority
down with yellow
warning symbol
at 0.7 Hz with
50% usage
symbol at 1.4 Ghz
with 50% usage
Audible Alarm
(Medium) 1 Burst with
3 pulses each, repeat
every 20 seconds
(high) 1 Burst with
10 pulses each, repeat
every 7.5 seconds
1 Burst with 3 pulses
each, repeat every
20 seconds
1 Burst with
10 pulses each, repeat
every 7.5 seconds
33