by the manufacturer for level II service every 2 years.
Manufacturer full service
The device shall be returned to O-Two™ Medical Technologies for manufacturer
full Service every 6 years.
Description
Visual
PM
inspection
Level II
Servicing
service
Servicing
Full service
Note: Upon request, the manufacturer provides access to factory-trained service
personnel and essential technical information. This includes circuit diagrams,
component part lists, detailed descriptions, calibration instructions, and any
other information necessary for the service and repair of the device.
Note: Single-patient use circuits shall be disposed of under local protocols.
At the end of their expected service life, the device, power supply, cables, and
battery shall be disposed of following local protocols.
38
Procedure
User Manual
Chapter 8.3
order, gas tanks are full,
Monthly
checking
Service manual
Service manual
Criterion
Device in working
no missing item
Meet product
specifi cations
Meet product
specifi cations
Schedule
Factor:
Monthly
User
Every
Manufacturer/
2 years
service center
Every
Manufacturer
6 years
eAdvantage User Manual