Download Print this page
O-Two e600 User Manual

O-Two e600 User Manual

Automatic transport ventilator

Advertisement

Quick Links

e600
Automatic Transport Ventilator
User Manual
e600 User Manual
(15PL1010 - Rev.7 28_8_2017)
English version

Advertisement

loading

Summary of Contents for O-Two e600

  • Page 1 Automatic Transport Ventilator User Manual e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 2 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 3: Table Of Contents

    Startup & setting the ventilation parameters ………..……….. Ventilation modes……………………………………………………….…………. 5.2.1 A/C V (Assist Control Ventilation) …………………………………. 5.2.2 SIMV (Synchronized Intermittent Mandatory Ventilation) ……………………………………………………………….…..……. 5.2.3 CPAP (Continuous Positive Airway Pressure)………….….. 5.2.4 CPR mode …………………………………………………….………….…. Turning Ventilator OFF ……………………………………………………….. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 4 9.6 Oxygen consumption …………………………………………………………….. 9.7 Oxygen Concentration delivered against different back pressures ………………………………………………………………………..……….. Trouble Shooting ……………………………………………………………….…………. Chapter 10 Abbreviations and Acronyms………………………………………….…………… Chapter 11 Chapter 12 Accessories ……………………………………………………………….…………………… Warranty ……………………………………………………………………………….……….. Chapter 13 e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 5: Safety

     Only use O-Two specified hoses, patient circuits and external power supplies to avoid affecting the output performance of the ventilator. Antistatic or conductive hoses or tubing are not used in the Ventilator Breathing System.
  • Page 6  The ventilator is intended for use in the prehospital, intrahospital, interhospital ground transport environments only.    Connecting the patient circuit to a patient prior to turning on the ventilator may cause a calibration error. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 7 .    Spare O-Two patient circuits are recommended at all times.    The design of this Ventilator does not incorporate a negative pressure phase during either automatic or manual ventilation.
  • Page 8: Chapter 2 Intended Use

    Chapter 2 Intended Use The e600 is a time-cycled and volume-constant emergency and transport ventilator designed for use in the pre-hospital, intra-hospital, inter-hospital and transport settings. It is intended for use with adult, child, infant patients with a tidal volume from 50 ml upwards who are in respiratory and/or cardiac arrest or respiratory distress and who require the ventilatory support.
  • Page 9: Chapter 3 Overview

    K – Cancel button E – Manual / Hold button L – Confirmation indicator light F – Screen brightness button M – Control Selection Knob G – Alarm Silence button N – ON/OFF button e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 10: Function Keys

    To lock the key membrane or cancel the lock function: 1- Press Lock button (I) in Figure 1. The Lock symbol will be displayed on the screen. 2- To cancel the Lock function, press the Lock button (I) again. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 11: Alarm Silence

    3.2.8 Pause / Resume During activation of the Pause/ Resume button (H) in Figure 1, the ventilator will stop ventilating with all buttons kept active (if they are not locked) except Manual/ Hold button. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 12: Manual / Inspiratory Hold

    The maximum inspiratory hold time is 6 seconds. After that time, the ventilator will switch to exhalation phase by opening airway pressure to ambient. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 15: Screen Layout

    SIMV screen layout CPAP screen layout CPR screen layout e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 16: Live Monitoring Parameters

    CPAP mode. This number will be updated at the end of each inspiratory phase. Note Mve, Rate and Paw are not active during CPR mode and displayed with “--“ e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 17: Symbols & Notations

    Ingress protection rating: Splash-proof. Do not immerse. Class II equipment. Protection against electric doesn’t rely on Basic Insulation only, additional safety precautions such as Double or Reinforced insulation are provided Type BF applied part. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 19: Connecting The Gas Supply

      Warning Extra care must be taken when handling oxygen: o The e600 must only be used with medical oxygen. o Only use approved medical oxygen compressed gas cylinders. o Always begin use with a full oxygen cylinder. o Secure oxygen cylinders so they do not fall over.
  • Page 20: Connecting Patient Circuit

    Visually inspect the ventilator for mechanical damage 2 Ensure that battery is fully charged. 3 Ensure the e600 is connected to a gas supply (Cylinder or piping system) that delivers a minimum of 45 PSI (3 Bar) and a maximum 87 PSI (6 Bar) output pressure and a minimum 120 L/min flow  ...
  • Page 21 If leak is still present and no external leak was detected using the above processes, the unit must be returned to the manufacturer or its authorised service center for service or repair. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 22 13- Turn off oxygen cylinder and check for low pressure and no pressure audible and visual alarms. 14- Turn off Ventilator by pressing ON/OFF button for 4 seconds. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 23: Chapter 5 Operating Instructions

    Control Selection Knob (M) to start ventilation. If no selection occurs within 20 seconds, the ventilator will start ventilation with child setting as its default start-up. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 24 If no selection occurs within 10 seconds or the Control Selection Knob (M) is not pressed to confirm the changed parameter setting within 10 seconds, changes will be cancelled and the previous parameter values will remain. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 25 0.66 1.33 100% 100% 100% F trig. (L/min) A/C V mode default settings Table-1 Figure-7 A/C V mode screen w/ default parameters (Child setting) Showing frame around Parameter during selection Figure-8 e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 26: Ventilation Modes

    Figure-9 5.2 Ventilation Modes The e600 ventilator is equipped with a number of ventilation modes to enable the healthcare provider to tailor the ventilator settings to the patient’s specific respiratory requirements. Ventilation could be delivered invasively (ET tube) or non-invasively (mask).
  • Page 27 O during I time only Manual Refer to Manual and I-Hold section ready ready ready * Ti may be limited below its range depending on set I:E ratio and rate. Figure-10 A/C V waveform e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 28: Simv

    0-20 cm H (during I time only) Refer to Manual and I-Hold Manual ready ready ready section * Ti may be limited below its range depending on set I:E ratio and rate. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 29 Figure-11 SIMV waveform e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 30: Cpap (Continuous Positive Airway Pressure)

    Mv (A) Will be calculated based on Vt & f I:E ratio (A) (1:4 – 3:1) P max. 10 -80 cm H P min. 0-20 cm H O during I time only e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 31 Figure-12 CPAP ventilation waveform e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 32: Cpr Mode

    Table-5 Default Ventilation Setting- CPR Parameter Range Default Infant Child Adult Tidal Volume (50 - 1400 ml) P max. (10 -80 cm H P min. (0-20 cm H O during I time only) e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 33 CPR for masked patients On screen chest compressing animation On screen ventilation animation Figure-13.a CPR waveform for masked patient e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 34 CPR for intubated patients On screen intubated CPR animation Figure-13.b CPR waveform for intubated patient e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 35: Turning Ventilator Off

    5.3 Turning the Ventilator Off Press and hold the ON/OFF button for 4 seconds, the Ventilator will turn OFF. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 36: Chapter 6 Post Use

    Unplug the power cable from mains if no charging is required. Clean and disinfect according to section 8.1 in this manual 6.2 Storage Store ventilator within the following environmental range: - 40 C to +60 C, Rh: 15% to 95%. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 37: Chapter 7 Alarms And Indicators

    In this case the visible and audible alarms will be based on highest priority alarm. All ventilation alarms of e600 are listed in the following Table-6. e600 User Manual (15PL1010 - Rev.7 28_8_2017)
  • Page 38 Solid symbol Solid symbol Patient effort during Patient effort Solid symbol Invalid setting- Refer During invalid to manual selection Solid symbol Setting Conflict During invalid selection Flash symbol Confirm after primary selection e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 39: Battery Status Indicator

    Fully charged Batteries must be recharged after 6 months without usage or when battery discharging LED turns orange and flashing. Battery Re-charging time is about 5.5 hours from fully discharged. Batteries have a minimum of 200 discharge and charge cycles. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 40: Leds

    ON and OFF phases. During OFF phase this light will start flashing when battery capacity drops to around 90 %. Green color LED - Continuous when unit is operated using internal battery. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 41: Chapter 8 Cleaning, Preventive Maintenance & Servicing

    Do not use chlorine based cleaning agents. Make sure no liquids enter the ventilator connections or the ventilator. Do not immerse the e600 Ventilator or patient circuit or supply hoses in disinfectant or other liquids, serious electric shock hazard and damage to the ventilator may occur.
  • Page 42: Ambient Air Entrainment Filter

    5 hours on a single charge. Ambient Air Entrainment Filter The e600 entrains ambient air through the internal Venturi system for ventilation when the O concentration is set at 60%. This provides not only decreased oxygen concentration but also increases the ventilator operating time on an oxygen cylinder.
  • Page 43 The device shall be returned to the manufacturer or a service center authorized by the manufacturer for bi-annual Level II service every 2 years. Manufacturer full service The device shall be returned to O-Two Medical Technologies for Manufacturer Full Service every 6 years. Warning...
  • Page 44: Chapter 9 Technical Data

    4000m (13000 feet) Battery Hot Swap Built-in Battery charger AC/DC power supply 100-240 VAC/ 19 VDC, 4.74 A Patient circuit O-Two Electronic Ventilator Circuit Mounting brackets for road ambulance and Mounting Bracket mobile setting Display 4.3" Color TFT...
  • Page 45 * BTPS: Volume measurements corrected to Body temperature 37 C and Barometric pressure 101.3Kpa under saturated conditions (100% Humidity). Note Measurement uncertainty: 5% for volume parameters and 6% for pressure parameters. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 47: Battery And Power Supply

    642 g EMC performance Meet requirements of IEC60601-1-2 Meet requirements of IEC60601- Safety Standards 1:2005 Note Upon disconnecting AC Power supply, Ventilator will automatically switch to Battery operation without affecting ventilator behaviour. e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 49 30 V/m IEC61000-4-3 80 MHz to 2.5 GHz O-Two e600 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the O-...
  • Page 51: Trouble Shooting

    Charge Battery as per discharges charging/Faulty instructions/Replace battery quickly battery Battery empty/no Change Battery/connect power e600 cannot power supply supply/ send to O-Two for repair or be switched on connected/defective service e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 52: Abbreviations And Acronyms

    Maximum airway pressure Minimum airway pressure SIMV Synchronized Intermittent Mandatory Ventilation Apnea Alarm time APNEA Expiratory time Thin Film Transistor Inspiration time Trig. Triggering flow rate Exhaled Tidal volume Tidal volume e600 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 53: Accessories

    Chapter 12 Accessories Item Part Order number O-Two Single-Use Electronic Transport Ventilator 01CV8030 circuit 6’ (1.83 m)patient circuit PVC 6’ (1.83 m) input pressure hose – 9/16” DISS nut 01FV4302 * fitting Intake filter-case of 10 01CV8040-cs Power supply Cord (Canada and the US)
  • Page 54: Warranty

    O-Two Medical Technologies Inc. cannot be held responsible for damage caused by non-compliance with the recommendations given above. The warranty and liability provisions of the terms of sale and delivery of O-Two Medical Technologies Inc. are likewise not modified by the recommendations given above.
  • Page 55 User Manual (15PL1010 - Rev.7 28_8_2017) English version...
  • Page 56 SN ____________ O-Two Medical Technologies Inc. 45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4 Office: +1 905 792 OTWO (6896) N.A. Toll Free: +1 800 387 3405 Fax: +1 905 799 1339 Email: resuscitation@otwo.com Web: otwo.com Marcel Houben Rue Vinave, 32...

This manual is also suitable for:

E series