1. SAFETY

WARNINGS
Federal Law restricts this device to sale by or on the order of a physician.
The eAdvantage® should only be utilized by qualifi ed healthcare
professionals who have received training as specifi ed in the 'Intended
Use' section.
The eAdvantage® system is designed for patient self-administration,
intended to be used under the supervision of a healthcare provider.Never
attach the face mask to the patient using a head harness.
The eAdvantage® shall not be used in oxygen-rich, fl ammable, or
anesthetic agents environments. Keep away from open fl ames, sparks,
and grease/oil.
It is recommended to use cylinders that are at least 1/4 full. Always turn
ON the cylinder valve slowly and fully.
The nitrous oxide cylinder should be operated in an upright position.
IOperating it in a valve-down position can result in liquid nitrous oxide
being expelled, which can cause freeze burns upon contact with skin.
Always use the checklist to ensure the correct assembly of all components
and replacement of disposable items.
Always use the checklist to ensure all components are reassembled
correctly and all disposable items are replaced.
Before use on a patient, the settings of the delivered gas should be
checked to ensure they are for the intended use.
It is recommended to use an oxygen monitor complying with ISO 80601-
2-55 whenever the gas mixer is in use by connecting it to the gas outlet
on the device between the device and the circuit.
This device must be used with the O-Two™ Breathing Circuit, 01CV8037.
It is recommended to connect the expiratory tubing of the circuit to an
Anaesthetic Gas Scavenging System complying with ISO 80601-2-13
If the alarm sounds continuously, immediately discontinue use and shut
OFF the gas supply.
After use, always turn OFF the cylinders and ensure that they have
su cient gas volume before returning the unit to its normal storage
position.
After use always turn OFF the device.
Do not disassemble any part of the unit except where described in this
manual as any unauthorized disassembly will invalidate the warranty.
Never obstruct the output port or the scavenging connector outlet.
DO NOT use the Oxygen Flush function to provide positive pressure
ventilation to a non-breathing patient.
4
eAdvantage User Manual