O-Two eAdvantage User Manual page 17

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Table of Contents

Advertisement

Once the operation screen is displayed with all the relevant parameters, the
eAdvantage® mixer is ready for operation.
Note: Start-Up Calibration Error (Startup calibration results and during use
technical failure):
The eAdvantage® will not switch ON if any of the problems listed below are
detected:
1.
O
Valve Failure.
2
2.
N
O Valve Failure.
2
3.
O
Flow Sensor Failure.
2
4.
N
O Flow Sensor Failure.
2
5.
U12 (Paw) Sensor Failure.
6.
U13 (O
7.
U14 (N
Instead, a new screen will display the cause of the malfunction and request the
mixer to be checked or returned for service. At the same time, a continuous
Audible Alarm will sound.
The above failures will be saved in the events log under specifi c failure as listed
in the "Stored History" section.
If multiple failures are detected the mixer will switch between the messages
related to the failures.
eAdvantage User Manual
input pressure) Sensor Failure.
2
O input pressure) Sensor Failure.
2
17