8.2 Charging the Battery

8.2.1 To charge the battery, connect one end of the external power supply/
charger to a power source (ranging from 100 to 240 Volts or an onboard vehicle
socket). Then, connect the other end to the DC input socket (C), as shown in
Fig. 1 on the side panel of the mixer. Upon successful connection, the indicator
light will activate, indicating the beginning of charging.
8.2.2 Turn the unit ON and observe the battery level (section 1 of the screen).
Refer to the 7.3 battery status indicator for exact battery charging status. The
battery shall be fully charged.
Note:
Battery Life expectancy:
Given normal storage and usage, the battery delivers 75% more of its initial
capacity after 300 charge/discharge cycles where the charge phase is CC/CV
3.3A, 16.80 V and the discharge is 3,3A down to 9000-mV pack voltage at 25
Battery Shelf life:
The battery provides a minimum of 6 months shelf-life with its initial charge
state of 40% when stored at 25
The battery pack should be replaced after its life expectancy. The same goes if the
battery pack will not fully charge (as indicated on the battery display on the mixer) or
if the mixer doesn't run for more than 5 hours on a single charge.
36
WARNING
The ambient temperature shall be between
0
C and 35
C during battery charge.
o
o
C.
o
WARNING
C.
o
eAdvantage User Manual