8. CLEANING, PREVENTIVE MAINTENANCE
AND SERVICING

8.1 Cleaning

Routine cleaning of the device shall be undertaken to maintain the device in a
clean condition.
The patient circuit, including the mouthpiece or facemask, is designed for
single-patient use only. Following each patient use, it should be discarded as
per local protocols and replaced with a new circuit.
Other components of the device should be wiped clean using a mild soap
solution or a hard surface disinfectant suitable for the device's materials. It is
critically important to ensure that the entire unit is never soaked or immersed
in any cleaning solutions under any circumstances.
Detailed cleaning procedures are as follows:
8.1.1 Ensure that the device is turned OFF and disconnected from the gas
supply source.
8.1.2 Remove the single patient-use circuit from the device and dispose of
safely following local protocols.
8.1.3 Wipe clean the N
O and O
input hoses with a mild soap or hard surface
2
2
disinfectant. Ensure no cleaning solution enters the hoses.
8.1.4 The enclosure of the device can be wiped over with a soft cloth and mild
soap solution or hard surface disinfectant. Ensure no cleaning solution enters
the input fi ttings.
8.1.5 If there is ingrained contamination a soft bristled brush may be used.
8.1.6 Dry all components thoroughly.
8.1.7 Attach a new patient circuit and connect the unit to the gas supplies prior
to use with the next patient.
WARNING
Do not attempt to clean and sterilize any components
that are designated for single-patient use.
35
eAdvantage User Manual