Cleaning, Preventive Maintenance And Servicing; Cleaning - O-Two eAdvantage 01EQ3000 User Manual

Electronic nitrous oxide/oxygen analgesic gas mixing and delivering system
Hide thumbs Also See for eAdvantage 01EQ3000:
Table of Contents

Advertisement

8. CLEANING, PREVENTIVE MAINTENANCE
AND SERVICING

8.1 Cleaning

Routine cleaning of the device shall be undertaken to maintain the
device in a clean condition.
The patient circuit with mouth piece (or facemask) is intended for
single patient use only and shall be discarded after each patient (in
accordance with local protocols) and replaced with a new circuit.
All other components should be wiped clean with a mild soap solution
or hard surface disinfectant suitable for the materials of manufacture
of the device. Under no circumstances should the complete unit be
allowed to be soaked or immersed in cleaning solutions.
Detailed cleaning procedures are as follows:
8.1.1 Ensure that the device is turned OFF and disconnected from the
gas supply source.
8.1.2 Remove the single patient use circuit from the device and dispose
of safely in accordance with local protocols.
8.1.3 Wipe clean the N
O and O
input hoses with a mild soap or hard
2
2
surface disinfectant. Ensure no cleaning solution enters the hoses.
8.1.4 The enclosure of the device can be wiped over with a soft cloth
and mild soap solution or hard surface disinfectant. Ensure no cleaning
solution enters the input fi ttings.
8.1.5 If there is ingrained contamination a soft bristled brush may be
used.
8.1.6 Dry all components thoroughly.
8.1.7 Attach a new patient circuit and connect the unit to the gas
supplies prior to use with the next patient.
WARNING
Do not attempt to clean and sterilize any components
that are designated as single patient use.
35
eAdvantage User Manual (15PL1033 Rev. 7, July 2022)