ERROR MESSAGES

Message
Aborted
Air Disabled
A Dose Error
Bas Disabled
Bas Sen Err
Basin Temp
B Dose Error
Clear Sys Alarm?
Flash High Error
Flash Low Error
Flo Disabled
Flow Sen Err
Flow 2 Sen Err
HLD Sampl Er
Lid ajar
Lid Disabled
Low Alcohol
Low Chamber
Description
Indicates a cycle was manually aborted and not completed
Indicates that the air flow sensor was disabled
Alarm indicating that the SSG Part A dosing reservoir has timed out and not all
chemistry was delivered
Indicates that the basin level sensor was disabled
Indicates that the basin level sensor is reporting a full basin at the start of a cycle (basin
should be empty)
Indicates that the basin temperature has not reached the programmed minimum
Alarm indicating that the SSG Part B dosing reservoir has timed out and not all
chemistry was delivered
Diagnostic 70: prompts the user to press the Enter button to clear a system alarm
(such as a NVRAM error)
Indicates that the high code flash CRC test did not pass
Indicates that the low code flash CRC test did not pass
Indicates that the fluid flow sensor was disabled
Indicates that the flow sensor is reporting flow at the start of a cycle (no flow should be
occurring)
(SSG only) alarm indicating that the chamber fluid flow sensor is reporting flow at the
start of the cycle (no flow should be occurring)
Alarm indicating that a disinfectant sample port sensor is reading incorrectly
Indicates that the cover (lid) is open during a cycle
Indicates that the cover (lid) open sensor was disabled
Indicates that the alcohol level is low
Indicates that the basin level did not reach the level sensor in time