Quick Links

Product Bulletin
Date: November 2, 2015
RE: SCOPE BUDDY
This bulletin provides alternative instructions for decontamination using chlorine tablets in markets where
these products are available.
The Daily Decontamination Procedure must be performed at the start of the day before use on any
endoscopes. (Refer to User Manual for more detailed instructions)
To ensure correct SCOPE BUDDY Endoscope Flushing Aid operation, the Quality Assurance Flow Validation
Test must be performed prior to the first use of the day.
Contact your local Medivators Customer Support for further assistance.
Prior to First Use of the Day
Perform Daily Decontamination Procedure by:
1. Locate Fluid Intake Line and place filter screen in decontamination solution in jug (Milton/Peracetic
Acid/Chlorine Tablets) Use concentration as per manufacturers guidelines.
2. Locate open end of outflow tubing and place in the decontamination solution which has been made up
in an empty jug/container (Milton/Peracetic Acid/Chlorine Tablets).
3. Using the front panel arrow keys, set the timer for 30 seconds.
4. Press "Start/Stop" button to begin fluid flow.
5. Prime the tubing until a steady column of decontamination solution exits the outflow tubing.
6. Press "Start/Stop" button to halt fluid flow.
7. Allow decontamination chemical to remain within Scope Buddy and tubing as per chemical manufacturer's
labelling for the appropriate contact time required to achieve decontamination.
8. Remove outflow tubing from jug and press start to remove decontamination solution from tubing by
purging lines with air.
9. Fill jug/container with clean rinse water and place intake line in jug.
MEDIVATORS
and SCOPE BUDDY
®
Endoscope Flushing Aid Daily Quality Assurance
®
are registered trademarks of Medivators Inc.
®
Medivators Inc.
14605 28th Avenue North
Minneapolis, MN
55447-4822 USA
Tel: +1.800.444.4729
Fax: +763.553.3387
www.medivators.com
For Distribution To:
• International Distributors
• International Management
• U.S. Field Sales & Clinical Specialists
• U.S. Technical & Field Service
Medivators BV
Cantel Medical Asia/Pacific Pte. Ltd.
Sourethweg 11
1 A International Business Park
6422PC Heerlen
#05-01 Singapore 609933
The Netherlands
Tel: +65.6227.9698
Tel: +31.45.5.471.471
Fax: +65.6225.6848
Fax: +31.45.5.429.595
Cantel Medical Devices (China) Co. Ltd.
Unit 804-805, Innov Tower Block A,
Hongmei Road, Xuhui 200233 Shanghai
Tel: +86 21 60161380
Fax: +86 21 61210913
50098-974-EN Rev A

Advertisement

loading

Summary of Contents for Cantel Medical MEDIVATORS SCOPE BUDDY

  • Page 1 Medivators Inc. ® ® Medivators Inc. Medivators BV Cantel Medical Asia/Pacific Pte. Ltd. Cantel Medical Devices (China) Co. Ltd. 14605 28th Avenue North Sourethweg 11 1 A International Business Park Unit 804-805, Innov Tower Block A, Minneapolis, MN...
  • Page 2 Product Bulletin 10. Place outflow tubing in sink 11. Using the front panel arrow keys, set the timer for 30 seconds 12. Press “Start/Stop” button to begin fluid flow 13. If SCOPE BUDDY Endoscope Flushing Aid is to be left unused for an extended period of time (i.e. ®...