Error And Warning Messages - Cantel Medical MEDIVATORS DSD EDGE Service Manual

Endoscope reprocessing system
Table of Contents

Advertisement

Error and Warning Messages

Error messages are displayed on the LCD screen to alert the operator to operational malfunctions and/or operational warnings (see
the Appendix for message definitions). If none of the solutions correct the problem, or if the problem recurs, contact your customer
support.
"Bas Sen Err" is displayed.
Incoming water pressure too high........................
Fluids on basin sensor.........................................
Basin fluid did not drain .....................................
Drain line kinked ................................................
"Flow Sen Err" is displayed during initial start.
Flow sensor stuck in the "on" position ...............
"Lid Ajar" is displayed during initial start.
The reprocessor lid is open .................................
"Low Chamber" is displayed during disinfectant phase.
Disinfectant screen is plugged..............................
Insufficient air pressure.......................................
Air pressure switch harness disconnected...........
Debris in the filter connections...........................
"Low Chamber" is displayed during disinfectant phase or rinse phase.
Insufficient air pressure.......................................
Insufficient water supply pressure, flow rate......
External water pre-filter plugged........................
Internal water filter is plugged............................
Incoming water regulator setting is incorrect .....
Air pressure switch harness disconnected...........
"No Air Flow" is displayed.
Air compressor is not working............................
Air filter is blocked .............................................
Hookup is disconnected or kinked ....................
Air filter is disconnected.....................................
Insufficient external air pressure.........................
126
Verify water pressure is set at 35-40 PSI.
Clean fluid droplets off sensor.
Contact your customer support.
Check the line for kinks. Reposition, if necessary.
Press the STOP button to clear the error.
Press the ADD AIR button to free the sensor.
Close the reprocessor lid. Verify there are no obstructions
preventing the lid from completely closing. Press the START
button to resume the cycle.
Check the screen. Clean, if necessary.
Check that the compressor is working correctly.
Check the external air regulator on non-compressor reprocessors.
Contact your location building maintenance.
Check the harness for proper connection. Also, check for
damage, loose wires.
Remove the filter and check the connections for debris. Clean
the connections and replace the filter.
Check that the compressor is working correctly. Check the
external air regulator on non-compressor reprocessors.
Contact your location building maintenance.
Contact your location building maintenance.
Replace the filter cartridge.
Replace the water filter.
Check the regulator setting. The setting must be 35 to 40psi. Do
not adjust the regulator if the setting is correct—check for a plugged filter.
Check the harness for proper connection. Also, check for
damage, loose wires.
Contact your customer support.
Replace the filter.
Verify the hookup is not kinked and reconnect.
Verify the air filter is connected.
Contact your customer support.