11. "Alcohol" displays on the screen. To set the alcohol air purge time, do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
12. "Alcohol Inject" displays on the screen. To set the alcohol inject time, type the two-digit number of seconds,
then press ENTER.
The volume of alcohol is controlled by the number of seconds entered on the screen, up to a maximum of
59 seconds.
1 second injection
13. "Air" displays on the screen. To set the air cycle time, do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
14. The custom program setting is complete. Record the settings on the next page (Custom Program Reference Chart).
= 3 ml alcohol
|
57