1.
On the control panel, press SETUP.
2.
Type 41 on the keypad, then press ENTER.
3.
"Water temp test?" displays on the LCD screen.
Press ENTER to initiate water flow.
4.
As water flows into the basin, verify the dynamic
water pressure.
Dynamic water pressure at the inlet to the
reprocessor must be 35-40 psi (2.4-2.75 BAR).
5.
As water flows into the basin, verify the incoming
water temperature readings on the control panel
LCD screen. The incoming water temperature for
each basin displays in the second column.
Incoming water temperature should be
approximately 35°C (± 2°C).
Once the water pressure and temperature have been verified, press STOP.
6.
Water continues to flow until manually stopped or if the temperature reaches 40°C.
7.
"Basin drained?" displays on the screen. Once the basin is completely drained, press STOP to close the
drain valve.
Note: The basin drain valves close automatically after 250 seconds.
Figure 4.7 - Verify Dynamic Water Pressure
Figure 4.8 - Display Water Temperature
|
33