INSTALLATION AND MAINTENANCE

Proper maintenance will ensure effective disinfection and prolong the life of the reprocessor:
The disinfectant immersion period (contact time) is fixed at five (5) minutes.
All pressure regulators are factory-preset. Do not adjust the settings. Contact your Technical Support
representative for assistance.
Do not use alcohol or alcohol-based products to clean the reprocessor cabinet.
The hookups are not autoclavable and must be reprocessed by low-temperature disinfection only.
Replacement parts must be ordered from the manufacturer to maintain the warranty of the unit and/or part and
to ensure proper operation for reprocessing.
On a monthly basis (or if a leak is detected), inspect the reprocessor for basin damage or pipe and tubing
damage, which may result in leaks.
Only properly trained individuals may operate or service the reprocessor.

WATER QUALITY AND FILTRATION

Water supplied to the machine must be Potable with a dynamic pressure of 35-90 psi, minimum flow rate of 3.2
GPM and temperature of 95°F ± 4°F (35°C ± 2°C)
Ensure water hardness is less than 12gpg (200ppm) for optimal performance. Excessive water hardness may
void the warranty.
The high performance 0.2-micron bacteria-retentive filter included with the reprocessor is a bacteria-retentive filter.
The routine maintenance schedule recommends replacing the 0.2-micron bacteria-retentive filter every 6 months
or sooner, depending on the pre-filtration system and the quality of the incoming water.
The routine maintenance schedule recommends replacing the 1.0 micron and the 0.45 micron filter every
6 months or sooner, depending on the quality of the incoming water.
The incoming water supply must have a shut off valve (upstream of the external pre-filtration system).
The incoming water supply must be shut off at the end of every work day. Ensure that this water supply is
turned on prior to operating the reprocessor.
The reprocessor is supplied with a 2-stage pre-filtration system.
The pre-filtration system should be monitored for decrease in pressure, indicating blocked filter membranes.
Note: The DSD EDGE™ Reprocessor does not require backflow protection for the incoming
water supply, nor is a backflow protection device installed in the reprocessor. If building or
plumbing codes require backflow protection, then please consult your local codes, standards
and/or guidelines, for applicable requirements.
|