REPROCESSING CYCLE

STARTUP
FLUSH
WASH
(Optional)
DISINFECT
RINSE 1
If monitored sensors are not satisfied during startup, the cycle is halted until the
issue is resolved or cycle is cancelled.
Automated Leak Tester (Optional). If enabled, a 40 second leak test is
performed before starting the Flush phase.
Fresh water is flushed through scope channels and basin.
Detergent (Optional). If programmed, detergent is injected into the
channels prior to the fresh water flush.
Incoming water temperature is monitored to ensure water temperature is reaching
proper range for disinfection.
Concludes by purging scope channels with air and emptying the basin of water.
An additional one or two soak segments can be added to the Flush phase.
No detergent is injected during additional soak segments.
Endoscope channels are flushed with disinfectant.
Basin is filled with disinfectant solution and endoscope is soaked for the 5 minute
disinfectant soak time.
Concludes by draining disinfectant from the basin.
Basin is partially filled with fresh water as scope channels are flushed with fresh
water. Partially filled basin is then drained.
Basin is completely filled with fresh water as scope channels are flushed with
fresh water. Basin is then drained while the channels are purged with fresh water.
Concludes by purging scope channels with air.
|