1. On the control panel, press SETUP.
2. Type 5 on the keypad, then press ENTER.
3. "Program 1" displays on the screen. Type the number
of the program (1-9) on the keypad, then press ENTER.
4. "Flush" displays on the screen. To set the desired flush time,
do the following:
a. Type the two-digit number of minutes, then
press ENTER.
b. Type the two-digit number of seconds, then
press ENTER.
5. "Soak" displays on the screen. To set the desired soak time,
do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
6. "Soak Rinse" displays on the screen. To set the desired soak rinse time, do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
7. "Detergent Inject" displays on the screen. To set the desired detergent inject time, type the two-digit number of
seconds, then press ENTER.
The volume of detergent is controlled by the number of seconds entered on the screen, up to a maximum of
59 seconds.
1 second
8. "Rinse 1" displays on the screen. To set the desired primary rinse time (Rinse 1), do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
9. "Rinse 2" displays on the screen. To set the desired secondary rinse time (Rinse 2), do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
10. "Rinse 3" displays on the screen. To set the third rinse time (Rinse 3), do the following:
a. Type the two-digit number of minutes, then press ENTER.
b. Type the two-digit number of seconds, then press ENTER.
= 3 ml detergent solution
Figure 6.16 - Create a Custom Cycle Program