DISPLAY STATE TIME

A cycle is comprised of a number of states. Use the following procedure to view the state number and time for
both stations.
1. On the control panel, press SETUP.
2. Type 18 on the keypad, then press ENTER.
"State Time Display" displays on the screen.
3. View the current state number and time remaining for
each station.
4. To exit, press SETUP.

PRINT ENTIRE LOG

Print a copy of the disinfection cycle log, then clear the log, at regular intervals. Only the information saved since the
last time the log was cleared is printed.
1. Verify the printer is ON.
2. On the control panel, verify the correct station is selected.
If necessary, press STATION SELECT to choose the
station (A or B).
3. Press SETUP.
4. Type 21 on the keypad, then press ENTER. "Print Log?"
displays on the screen.
5. To print the log, press START.
6. To exit, press SETUP.
Figure 6.8 - Display State Time
Figure 6.9 - Print Entire Log
|