1. Connect the ports with the appropriate hookup.
For help on hookup selection, refer to the
Interactive Hookup Guide (Hookup Lookup) on
the Medivators website (www.medivators.com).
2. Do the following to correctly position the endoscope
in the basin:
a. Position the control section of the endoscope
in the right rear of the basin.
b. Position the light guide in the left front of
the basin.
c. Ensure the distal end is not pointing upward
toward the floating lid.
3. Attach the endoscope hookup connection to the
basin endoscope connection.
4. Verify there are no kinks in the hookup.
5. If the automated leak test function is enabled,
connect the leak test hookup to the basin leak
test connection.
Figure 5.1 - Position Endoscope in Reprocessor Basin
|
37