Display Log; Clear Log - Cantel Medical MEDIVATOR DSD EDGE User Manual

Automated endoscope reprocessor
Table of Contents

Advertisement

DISPLAY LOG

The entire log can be displayed one entry at a time, starting with
the most recent entry.
1. On the control panel, press SETUP.
2. Type 8 on the keypad, then press ENTER. The most recent
log entry displays.
3. Press ENTER to see the previous log entry.
4. To exit, press SETUP.

CLEAR LOG

The log stores 1463 records per station. Once the log is full, additional records will overwrite the oldest entries.
Print a copy of the log and clear the log at regular intervals.
Note: The selected station must be idle to perform this function.
1. On the control panel, verify the correct station is selected.
If necessary, press STATION SELECT to choose the
station (A or B).
2. Press SETUP.
3. Type 10 on the keypad, then press ENTER. "Clear Log?"
displays on the screen.
4. To clear the log, press ENTER.
To retain the log, exit without pressing ENTER.
5. To exit, press SETUP.
Figure 6.4 - Display Log
Figure 6.5 - Clear Log
|
49