Reprocessing Cycle Times - Cantel Medical MEDIVATORS CER OPTIMA User Manual

Endoscope reprocessing system
Table of Contents

Advertisement

Reprocessing Cycle Times

Note: All reprocessor models are capable of using EITHER the full automatic cycle
OR the HLD/rinse cycle.
Note: If using RAPICIDE
OPTIMA
an CIDEX
immediately following HLD step. CIDEX OPA requires these complete rinse cycles
after disinfectant has contacted the endoscope or medical devices.
"Full" Automatic Cycle
Step
Cycle
1
Wash
2
Drain/Air Purge
3
Rinse
4
Drain/Air Purge
5
HLD IN
6
HLD/Out/Air Purge
7
Rinse
8
Drain/Air Purge
9
Rinse
10
Drain/Air Purge
11
Air (if chosen)
12
Alcohol (if chosen)
HLD/Rinse
Automatic Cycle
"
"
Step
Cycle
1
Rinse
2
Drain/Air Purge
3
HLD IN
4
HLD/Out/Air Purge
5
Rinse
6
Drain/Air Purge
7
Rinse
8
Drain/Air Purge
9
Air (if chosen)
10
Alcohol (if chosen)
66
or Glutaraldehyde high-level disinfectants, then the CER
®
system will use 2 post rinses immediately following the HLD step. If using
®
OPA high-level disinfectant, then the CER OPTIMA will use 3 post rinses
®
Minimum Time (in minutes)
CER-1: 2, 3 or 5 minutes;
CER-2: 3 or 5 minutes
2.0
Depends on water flow
2.0
Depends on disinfectant type in use
2.0
Depends on water flow
2.0
Depends on water flow
2.0
3, 5 or 10
CER-1: 30, or 60 cc's;
CER-2: 30, 60 or 120 cc's
Minimum Time (in minutes)
1.0
1.0
Depends on disinfectant type in use
2.0
Depends on water flow
2.0
Depends on water flow
2.0
3, 5 or 10
CER-1: 30, or 60 cc's;
CER-2: 30, 60 or 120 cc's
Maximum Time (in minutes)
5.0
2.0
6.0
2.0
2.0
6.0
2.0
6.0
2.0
3, 5 or 10
3, 5 or minutes
(Depends on air settings)
Maximum Time (in minutes)
1.0
1.0
2.0
6.0
2.0
6.0
2.0
3, 5 or 10
3, 5 or 10 minutes
(Depends on air setting)