Renishaw TONiC T20 1 Series Installation Manual page 6

Rexm/rext ultra-high accuracy angle encoder system
Hide thumbs Also See for TONiC T20 1 Series:
Table of Contents

Advertisement

interface installation drawing
Dimensions and tolerances in mm
4- 40 UNC x 2
Cover plate
Diagnostic LED
( Ti0004 - Ti20KD and
TD4000 - TD0040 )
CAL button operation
Push and release (<3 seconds) - CAL routine enable/disable
Push and release (>3 seconds) - AGC enable/disable
Push and hold during power 'Off/On' cycle - Restore factory defaults
Refer to readhead LED functionality chart for CAL LED indications
8
40
16
33.3
62
67
6 min
R10 Static bend radius
R20 Dynamic bend radius
CAL/AGC push button switch
access hole Ø2.4
This section is a quick-start guide to installing a TONiC system.
More detailed information on installing the system is contained in the
following sections of the installation guide.
INSTALLATION
Ensure scale, readhead optical window and mounting faces are clean and free from obstructions.
If required, ensure reference mark selector magnet is correctly positioned.
Plug the readhead cable into the Ti interface under the cover plate and reassemble interface.
Connect to receiving electronics and power-up.
Ensure AGC is switched off - the CAL LED on the readhead should be off (if not press and hold the
CAL button on the interface until the CAL LED on the readhead switches off).
Install and align the readhead to maximise signal strength over the full axis travel as indicated by the
readhead and interface set-up LEDs (readhead green; interface - ideally blue/purple).
CALIBRATION
Press and release the CAL button on the interface.
The CAL LED on the readhead will be single flashing.
Move the readhead along the scale at slow speed (<100 mm/s), without passing a reference mark,
until the CAL LED starts double flashing.
If a reference mark is not being used, the calibration routine should now be exited
by pressing the CAL button.
Move the readhead back and forth over the selected reference mark
until the CAL LED stops flashing and remains 'off'.
The system is now calibrated and ready for use.
AGC can now be switched on if required by pressing and holding the CAL button
until the CAL LED on the readhead switches on.
CAL values and AGC status are stored in readhead non-volatile memory at power down.
NOTE: If calibration fails restore factory defaults by pressing and holding the CAL button
whilst switching on. Then repeat the installation and calibration routine.
quick-start guide