Download Print this page

HP 4277A Operation And Service Manual page 98

Lcz meter

Advertisement

SECTION m
Model 4277A
Table 3-13. Remote Program Codes (Sheet 1 of 2)
Item
Control
Program
Code
Description
DISPLAY A
Function
L
C
HIGH SPEED
HIGH SPEED
Z*
A1
A2*
A3
A4
A5
DISPLAY B
Function
D
Q
ESR/G
B1
B2
B3'
DISPLAY A and DISPLAY B com¬
binations are listed in the
table below:
B
A
1
2
3
1
L-D
L-Q
L-ESR/G
2
C-D
C-Q
C-ESR/G
* When DISPLAY A is set to
Z, DISPLAY B is auto¬
matically set to 6.
CKT MODE
AUTO
o-CS-Wr-o
C1'
C2
C3
MEAS SPEED
SLOW
MED
FAST
Ml
M2'
M3
Auto Range
OFF
ON
UO
:
Range is fixed.
Ul'
:
Range is automatically
selected.
LClZ
Range
lyH/lpF
R1
lOpH/lOpF
R2
lOOyH/lOOpF/lOQ
R3
ImH/lnF/lOOfi
R4
lOmH/lOnF/lkfi
R5
lOOrall/lOOnF/lOk^^
R6
IH/lyF/lOOkfi
R7
lOpF/lM^^
R8
If the instrument is set to
a range which cannot make the
measurement, range is auto¬
matically reset to the near¬
est range capable of making
the measurement.
Test Signal
Level
LOW
HIGH
VI
V2'
Trigger Mode
Execute
I NT
MAN/EXT
Tl'
This code only sets the trigger
T2
mode; it does not trigger the
instrument.
EX
This code is used to trigger
the instrument.
Self Test
OFF
SO'
ON
SI
Deviation
Measurement
OFF
ON
XO'
XI
Zero Offset
OPEN
SHORT
ZO
zs
3-56

Advertisement

loading