Model 4277A
12. Set the bias voltage to OV via the HP-IB.
OUTPUT 717; "BIOEN"
13. Wait until the dc bias voltage returns to OV.
* Example of programming a 10ms wait.
WAIT 10
SECTION m
14. Remove the DUT from the test fixture.
Note
The
above
remote
programming
examples can be used on the HP Model
85 (with 00085-15003 I/O ROM), Model
9835A, Model 9845B/C, Model 9826A,
and Model 9836A.
Note
In the above examples, HP-IB address
17 was used.
Figure 3-31. Option 001 Internal DC Bias (Sheet 3 of 3).