Download Print this page

HP 4277A Operation And Service Manual page 155

Lcz meter

Advertisement

Model 4277A
SECTION IV
Table 4-12. Controlier Instructions and Operator Responses for Test Program 1
Controller Instructions
Operator Response
Displays
Printout
***
4277A HP~IB TEST NO.l
***
REMOTE/LOCAL TEST
LISTEN=1*, TALK=0, REM0TE=1
LISTEN=0,
TALKED, REM0TE=1
L1STEN=0,
TALK=0, REMOTE=0
l:isten=i,
talk=o, remote=i
If the 4277A HP-IB
Status Indicators
and Controller
Display are the
same, press
0,
and
Fend
LiNel
If
not, press 3 ,
and
fiND linej .
PRESS LOCAL KEY
Press Local Key.
LISTEN=1, TALK=0, REM0TE=1
listen=i, talk=o, remote=
LISTEN=1, TALK=0, REM0TE=1
If the 4277A HP~IB
Status Indicators
and Controller
Display are the
same, press
0 ,
andlEND
LtNE'l.
If
not, press
[n] ,
and
flND line!.
REMOTE/LOCAL TEST PASS
REMOTE/LOCAL TEST PASS
If all steps are
correct, this
message is output.
REMOTE/LOCAL TEST FAIL
REMOTE/LOCAL TEST FAIL
If any step fails,
this message is
output.
LISTEN/TALK TEST
LISTEN-0, TALK=1, REMOTE-1
LISTEN=1,
TALK=0,
REMOTE-!
If the 4277A HP-IB
Status Indicators
and Controller
Display are the
same, press
0 ,
and
Fend uneI.
If
not, press
®
,
and
Fend uneI .
LISTEN/TALK TEST PASS
LISTEN/TALK TEST PASS
If both steps are
correct, this
message is output.
LISTEN/TALK TEST FAIL
LISTEN/TALK TEST FAIL
If any step fails,
this
message is
output.
END
END
*1 indicates ON; 0 indicates OFF.