Download Print this page

HP 4277A Operation And Service Manual page 20

Lcz meter

Advertisement

SECTION I
Model 4277A
Table l-L Specifications (Sheet 9 of 17)
Inductance Measurement Accuracy
L-D Measurement Accuracy 5
L Accuracy:
±[(% of reading) + (L error) + (number of counts)], see Tables R-1 and
B-2.
D Accuracy:
±[(% of reading) + (D error) + (number of counts)], see Tables B-1 and
B-2.
Note:
Use Table B-1 when the test frequency is lOkHz, lOOkHz, or IMHz,
Use Tables B-2 for all other frequencies.
Note:
Accuracies obtained from Tables B-1 and B-2 are valid only for
measurements made with the CABLE LENGTH switch set to Om.
When the CABLE LENGTH switch is set to Im, add the errors listed
in Table B-3 to the accuracies obtained from Tables R-1 and B-2.
Table B-1. L-D Accuracies (lOkHz, lOOkHz, IMHz only)
Inductance
Range
Test Frequency
lOkHz
lOOkHz
IMHs
IH
1
lOOrnH
1% + 5
lOmH
1
% + .02 + ^
k
Imll
.2% + .4yH + 5
.3% + .0005/a + .0008 +
S
,5% + 5
.5% + .005a + .005 + 5
k
lOOyH
..3% + 40nH + 10
.5% + .0005/a +
.005 + 5
.2% + 40nH +5
''
.3% + .0005/a + .0008 +
S
f
!
1
i
i
lOyH
,3% + 4nH + 10
.5% + .0005/a +
.005 * 5_
j-1
1
. 2% + 4nH + 5
1
|.3% + .0005/a + .0008 + 5_|
IpH
1
1
i.3% + .4nH + 10
1
1.5% + .0005/a + .005 + 5
|
1.
-I

Advertisement