Download Print this page

HP 4277A Operation And Service Manual page 152

Lcz meter

Advertisement

SECTION IV
Model 4277A
14. Check the comparison data output at the HANDLER INTERFACE connector by
comparing it with the Data States shown in Table 4-11.
15. Press the ERASE key and set the following comparator limits:
L/C/ I Z I
HIGH LIMIT
BINl:
.9
BIN2:
I.l
BINS:
1.9999
D/Q/ESR/G HIGH LIMIT:
.1
16. Press the RUN key on the 16064A's control panel.
17. Verify that the L/C/1 Z I
IN and D/Q/ESR/G
IN lamps light, and "2" is
displayed on the 4277A's FREQUENCY/DC BIAS DISPLAY.
L/C/IZI
o • o
D/Q/ESB/Q
LOW
IS
HIQM
o » o
FREQUENCY / DC BIAS
BIN NUWQER
18. Check the comparison data output at the HANDLER INTERFACE connector by
comparing it with the Data States shown in Table 4-11.
Table 4-11. Handler Interface Output Data States
TEST
Connector Pin Numbers
STEP
1 2 3 4 5 6 7 8 9 10
19 20 21 22 23 24
10
H 11 H H H H H H ii
L
H
11
L
L
H
H
13
H M H H 1! H H H H
L
L
H
H
11
H
L
17
H L M [■■! H H H H 11
H
H
L
11
H
L
H
H: Approximately 5V
L; Approximately OV
IB 17 16
3 2
I
36 35 34
2120 19
Figure 4-8. Handler Interface Connector Pin Assignments.
4-24