HP 4277A Operation And Service Manual page 123

Lcz meter

Advertisement

Model 4277A
SECTION III
* Example of setting bin 2*s low limit to
l.lOOlnF and high limit to 1.2nF
OUTPUT 717;"N2"
OUTPUT 7I7;"LL1.1001ENLH1.2EN"
Note
The same setting can be made by the
following program:
7. Enter the limits for D/Q/ESR/G via the
HP-IB.
* Example of setting a low limit of .0000
and a high limit of .005
10. Trigger the instrument via the HP-IB.
* Example of triggering the instrument:
OUTPUT 717;"EX''
If necessary, read the comparison results
via the HP-IB.
ENTER 717;A$
PRINT A$
Note
The HP-IB address code in the above
examples is 17 (10001).
OUTPUT 717;"L2"
OUTPUT 717;''LL0ENLH,005EN''
Note
The same setting can be made by the
following program :
OUTPUT 717;"L2"
OUTPUT 717;"LH.005EN"
Note
Comparator operations
can
be
done
without high/low limits for D/Q/ESR/G.
8. Start the comparator operation by HP-IB
program.
* Example
of
starting
the
comparator
operation:
OUTPUT 717;"G1"
9. Connect the DUT to the test fixture.
Figure 3-34. Option 002 Comparator (Sheet 5 of 7).