Download Print this page

HP 4277A Operation And Service Manual page 127

Lcz meter

Advertisement

Model 4277A
SECTION IV
Table 4-1. Recommended Equipment (Sheet 1 of 2)
Equipment
Critical Specifications
Recommended
Model/Note
Use*
Digital
Voltmeter
Voltage range:
lOmV to lOOVf.s.
Resolution:
O.lmV
Accuracy:
0.05%
Input impedance:
>10iyK2
HP 3478A
P, A, T
RF Voltmeter
Voltage range:
lOmV to 3Vrms f.s.
Bandwidth:
lOkliz to IMHz
Accuracy:
1%
HP 400E
and HP 3403C
P, A
Frequency-
Counter
Maximum frequency:
>lMHz
Accuracy:
0.001%
Trigger level:
Adjustable
HP 5314A
P,
A,
T
DC Power
Supply
Maximum output voltage:
>50V
Resolution:
<100mV
HP 6206B
P
Oscilloscope
Bandwidth:
lOOMHz
Sensitivity:
5mV/DIV
HP 1740A
A, T
Oscillator
Frequency:
Ikliz
Output voltage:
IraV
HP 652A
T
Signature
Analyzer
HP 5004A
T
Test Cables
BNC Cm)-to-BNC (m), 61cm long, 1 ea.
P, A
BNC (m)-to-BNC (m), 10cm long, 1 ea.
A, T
BNC (m)-to-BNC (m), 30cm long, 2 ea.
HP 11170A
T
BNC (m)-to-Duai Banana Plug, 1 ea.
HP liOOlA
P, A
Dual Banana Plug-to-Alligator Clip, 1 ea.
HP 11002A
P, A, T
BNC (m)-to-Dual Alligator Clip,10cm
long, 2 ea.
Refer to the
troubleshooting
diagram A2-17.
T
Alligator Clip-to-Alligator Clip,20cm
long, 1 ea.
HH
Adaptors
BNC (£)-tO"BNC (f), 5 ea.
HP P/N 1250-0080
Oscilloscope
Probes
10:1 Divider Probe
Input impedance:
10M^2
HP 10004D
A, T
1:1 probe
HP 10007B
A
Test Leads
HP 16048A
P, T
*P = Performance Test, A = Adjustment, T = Troubleshooting
4-A

Advertisement

loading