Download Print this page

HP 4277A Operation And Service Manual page 128

Lcz meter

Advertisement

SECTION IV
Model 4277A
Table 4-1. Recommended Equipment (Sheet 2 of 2)
Equipment
Critical Specifications
Recommended
Model/Note
Use*
Capacitance
Standards
lpF±0.03%
10pF±0,03%
100pF±0.03%
1000pP±0.03%
Useable frequency:
Up to IMHz
HP 16381A
HP 16382A
HP 16383A
HP 16384A
P, A, T
Resistance
Standards
lOQ.
i00fl±0.03%
lkj:!±0.03%
10kfi±0.03%
100kf2±0.03%
OPEN termination
SHORT termination
HP 16074A
Standard
Resistor Set
P, A, T
Capacitors
lnF±5%
HP P/N 0160-2218
T
Resistors
4.7,fi±5% 1/4W
HP P/N 0683-0475
T
560fl±5% 1/4W
HP P/N 0683-5615
T
lkfi±5% 1/2W
HP P/N 0757-0159
T
10kl^±l% 1/2W
HP P/N 0757-0839
T
100kSi±l% 1/8W
HP P/N 0757-0465
T
HP-IB
Controller
HP-85/
w 00085-15003/
w 82936A/
w 82937A
A
*P = Performance Test, A = Adjustment, T = Troubleshooting