HP 4277A Operation And Service Manual page 150

Lcz meter

Advertisement

SECTION IV
Model 4277A
4-25.
ieQ04A COMPARATOB/HANDLES^ mTEBEACB TEST (OPTION 002)
4-26. This test verifies the functions of the 16064A Comparator/Handler Interface.
Figure 4-7. 16064A Comparator/Handler Interface.
EQUIPMENT:
Digital Multimeter .HP 3478A
lOOkQ Standard resistor .HP 16074A
lOOOpF Standard capacitor .HP 16384A
PROCEDURE:
1. Set jumpers A1 Wi/W2/W3 in the 16064A to the lower position as shown below:
Upper Position
I6064A
Al
Boord
2. Connect the 16064A to the COMPARATOR/HANDLER INTERFACE connector
on the rear panel of the 4277A.
3. Turn on the 4277A. "16064" should be displayed on DISPLAY B.
4. Set the 4277A's controls as follows:
DISPLAY A/B functions .C-G
Test Frequency . l.OOkHz
DC BIAS .OFF
CKT MODE .
LC [ Z I
RANGE . InF
MEAS SPEED . MED
TEST SIG LEVEL. HIGH
TRIGGER .INT
4-22