SECTION I
Model 4277A
Table 1-2. Supplemental Performance Characteristics (Sheet 2 of 2)
Additional Measurement Error of Test Fixtures:
Maximum additional errors attributable to
the test fixtures:
Model
Residual Impedance
16047A, 16047C,
16048A, 16048B,
16048D, 16065A
-
16048C
C<App. 5pF
L<App. 200nH
R<App. lOmfl
16034B
C<App. 0.02pF
L<App. 30nH
R<App. 50mf2
Additional Measurement Error when D>0.01:
Add 5D% (when LOW test signal level is used,
10D%) to the accuracies for HIGH SPEED C
and HIGH SPEED L.
Measurement Time:
Typical
characteristics are
shown
in
the
figure below:
Additional Measurement Error when D > 0.1:
Multiply C, L, or D accuracy by (1 + D^)
Settling Time after measurement range change:
Approximately 60ms
Settling Time after frequency change:
Approximately 300ms
Settling Time after Test Signal Level Change:
Approximately 60ms
Test
Signal
Settling
Time
in
DC
Bias
applications:
The same as dc bias voltage settling time
DC Bias Voltage Settling Time:
Typical
value
for
C
measurement
(C <
2000pF)
Bias Voltage
Settling Time
99% of setting
7.5ms
99.9% of setting
25ms
99.99% of setting
40ms
Number
Measurement
Function
Measurement
Speed Mode
1
C
MED
2
C
FAST
3
L
MED
4
L
FAST
5
Z
MED
6
z
FAST
7
HIGH SPEED C-
HIGH SPEED L
MED
8
HIGH SPEED C
HIGH SPEED L
FAST
Continuous Memory:
Approximately 2 weeks
(at 23"C±5®C)

Advertisement