4-9.
TEST FREQUENCY ACCURACY TEST
4-10.
This test verifies that test signal frequencies for the 4277A meet the specified
frequency accuracy of 0.01%.
5314 A
4277A
* 1
1
••ry—,..";,
ill.
5
I
B
■ V C'C'
44'
i:-
i::.i B(B£3 c; a gicij
8NC to BNC
Coble
Figure 4-L Test Frequency Accuracy Test Setup.
EQUIPMENT:
Frequency Counter .HP 5314A
BNC to BNC cable .HP 11170A
PROCEDURE:
1. Connect the frequency counter to the 4277A UNKNOWN
Hcur
terminal as
shown in Figure 4-1.
2. Set the 4277A's controls as follows:
TEST SIG LEVEI..HIGH
DC BIAS switch.OFF
Test Frequency.lOkHz
Other controls.Any setting
3. Verify that the frequency reading on the 5314A is lO.OOOkHztlHz.
4. Set the test frequency in the sequence given in Table 4-2. Verify that the
frequency readings on the 5314A are within the test limits given in the table.
Table 4-2. Test Frequency Accuracy Test
Frequency Setting
Test Limits
lO.OkHz
9.999 to lO.OOlkHz
lOOkHz
99.99 to lOO.OlkHz
202kH2
201.98 to 202.02kHz
500kHz
499.95 to 500.05kHz
i.OOMHz
0.9999 to l.OOOlHHz
Note
1) Test limits in the table do not account for tolerance dependent on the
specified accuracy of the 5314A.
2) If this test fails, the instrument requires troubleshooting.