Download Print this page

HP 4277A Operation And Service Manual page 169

Lcz meter

Advertisement

Model 4277A
SECTION IV
PERFORMANCE TEST RECORD
Paragraph
Numb er
TEST
Results
Minimum
4-17
CAPACITANCE ACCURACY TEST (Cont'd)
lOOkHz
C
C.V.-O.OOlSnF
C.V.+0.0018nF
D
-0.0016
0.0016
200kHz
C
C.V."0.0023nF
C.V.+0.0023nF
0
-0.0026
0.0026
202kHz
c
C.V.-0.0027nF
C.V.-t-0.0027nF
D
-0.008
0.008
505kHz
C
C.V.-0.0021nF
C.V.+0.0021nF
D
-0.0022
0.0022
IMHz
C
C.V.-0.0015nF
C.V.+0.00i5pF
D
-0.0016
0.0016
Test Signal Level :
LOW
Frequency
lOkHz
C
C.V.-0.0036nF
C.V.+0.0036nF
D
-0.0032
0.0032
20kHz
C
C.V.- 0.014nF
C.V.+ 0.014nF
D
-0.014
0.014
20.2kHz
C
C.V.- 0.014nF
C.V.+ 0.014nF
D
-0.11
0.11
50.5kHz
C
C.V.- 0.013nF
C.V.^ 0.013nF
D
-0.013
0.013
lOOkllz
C
C.V.-0.0036nF
C.V.+0.0036nF
D
-0.0032
0.0032
200kHz
C
C.V.- 0.014nF
C.V.+ 0.014nF
D
-0.014
0.014
202kHz
C
C.V.~ 0.014nF
C.V.+ 0.014nF
D
-0.11
0.11
505kHz
C
C.V.- O.OlSnF
C.V.+ 0.013nF
D
-0.013
0.013
IMHz
C
C.V.-0.0020nF
C.V.+0.0020nF
D
-0.0032
0.0032
C.V. = Calibrated Value
IX