Xerox DocuColor 12 Color Manual page 53

Color server
Hide thumbs Also See for DocuColor 12:
Table of Contents

Advertisement

2
2-19 Using the EFI Spectrometer ES-1000 and EFI Densitometer ED-100
8. Choose Clicks Per Patch, Audio Feedback, and Port options, and then click Start.
The Page Type menu automatically chooses the measurement page type that you
printed in step 6.
Use the arrow buttons next to the Clicks Per Patch field to choose the number of
measurements you want to take for each patch on the page. When you choose multiple
clicks, the average value of all measurements taken for the patch is returned to the
Fiery. Multiple measurements are more accurate, but may take more time.
Select the Audio Feedback option if you want to hear a confirmation tone each time a
measurement is recorded. A single tone indicates a successful measurement; two tones
indicate a measurement error. If you have specified multiple clicks per patch, the tone
sounds after you take the final measurement.
If necessary, choose the appropriate COM port for the ES-1000 or ED-100 from the
Port menu.
When you click Start, the Measurements window appears. Directions for measuring
the patches appear in the Status area.
9. Place the ES-1000 or ED-100 over the specified patch on the Measurement Page,
making sure to center the sample aperture over the patch. Press Measure to take the
measurement.
For a more accurate measurement, place several plain sheets of paper beneath
N
:
OTE
the Measurement Page to block underlying colors from being read by the instrument.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fiery ex12Fiery ex2000d

Table of Contents