Measuring

Ensure that the Sample Measurement Surface and
1
Sample Well are visually clean, dry, and residue-free.
Refer to the Cleaning & Maintenance section below if
cleaning is needed.
Dispense the sample to cover the entire Sample
2
Measurement Surface (center black circle).
If necessary, wait for the temperature
3
to stabilize.
Press the Measure button to perform
4
the measurement.

Temperature Monitor

The temperature monitor displays the current
temperature of the Sample Well and a graph illustrating
how much the temperature is changing. If a hot or cold
sample is dispensed onto the sample well, that rate-
of-change will increase, and the temperature readout
will turn red to alert the user that the temperature is
unstable.
To ensure accurate results, it is recommended to
wait until the temperature readout is black, indicating
the temperature has stabilized, before taking a
measurement.
13910000-101 Rev. A • 20
4.9ºC
22.0ºC