Maintenance; Rectification Of Faults And Malfunctions; Faults Table - GARANT VG BASIC Instruction Handbook Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 33
GARANT VG Basic pre-setting device
de
9.

Maintenance

DANGER
en
Electrically live components
Risk of fatal electric shock.
»
Before starting any maintenance work switch the unit off and disconnect it from the power supply, release it and pro-
tect it from being switched on again.
cs
»
Check that the system is electrically dead, earth it and short circuit it.
»
Insulate any adjacent electrically live components.
»
Maintenance and repair may be performed only by electrical specialists.
da
»
If any live components suffer damage, have them repaired immediately.
Interval
Daily or as required
es
fi
Weekly or as required
fr
hr
hu
Annually or as required
it
lt
nl
According to the applicable national
regulations
pl

10. Rectification of faults and malfunctions

10.1.

FAULTS TABLE

Fault
ro
No camera image, or the
camera image is black.
ru
Device cannot be calibrated. No calibration master gauge
sl
sv
56
Maintenance work
To avoid corrosion, depending on
the working environment, daily
grease the tool mounting after use,
or rub a lubricant such as WD-40
into it.
Wipe down the housing with a
slightly damp cloth.
The optical components such as the
lens and camera should be kept free
of oil, dust and fingerprints. Clean
them with a non-linting cloth to
which a little alcohol has been ap-
plied.
Clean the screen with a commer-
cially-available glass cleaner and a
soft cloth.
1. Move the guided units to their re-
spective end positions.
2. Inspect the guides for the tower and
anti-rotation arm and grease them
as necessary.
»
Use multi-purpose grease. Re-
commendation: LGEP 2/1 from
SKF.
Check the calibration and check the
guides for play.
Check the electrical components (in-
cluding the power supply unit) for con-
formity to national regulations.
Possible cause
Camera software has crashed. Restart the panel PC.
Cable open circuit
Incorrect lighting setting.
in the live image
Performed by
Trained person
Trained person
Trained specialist for mechanical work
Action
Check the plug connections.
Check the lighting setting.
1. Move the adapter/calibra-
tion mandrel into the
measurement window.
2. Use the dynamic focus
display [} Page 49] to
move the adapter/calibra-
tion mandrel into focus.
Performed by
Trained person
Trained person