1. Using the dynamic focus display [} Page 49], rotate the tool/calibration mandrel to the largest displacement.
2. Press Next point to start the pixel calibration.
»
A green rectangle appears on the screen.
3. Using the infinite fine adjustment to move the calibration edge into the green rectangle.
4. Press Next point to continue the pixel calibration.
5. For each pixel, use the infinite fine adjustment to move the calibration edge into the green rectangle.
»
Pixel calibration ends automatically when the last pixel is reached (18/18).
i
After pixel calibration, the adapter currently in use must be recalibrated.
8.5.4.

Parallelism calibration

i
Parallelism calibration allows parallelism errors between the Z-axis and the rotary spindle/rotary axis to be corrected.
ü The date of the last parallelism calibration is shown in the measurement software.
ü Insert a plug gauge of sufficient length (300-500 mm) into the tool mounting.
1. Move the edge into the live image.
2. Move the camera to the bottom end of the mandrel and acknowledge this with
3. Move the camera to the top end of the mandrel.
»
When the axis has come to a halt, the calculated deviation and angular error are displayed in a window.
»
At every measurement, the software will incorporate this dimension into the calculation, and correct it.
4. After the parallelism calibration, save the new setting for the absolute zero point.
»
Changes to the basic setting will invalidate the calibration of the pre-setting device.
5. Recalibrate the adapter currently being used.
8.6.

DEVICE INFORMATION

i
Information about the version and serial number. The serial number can be found on the nameplate.
8.7.

EXIT

Press Exit and go to the Windows interface to set up the printer, for instance.
Press Exit and switch off to shut down the panel PC. Switch off at the main switch to fully shut down the device.
Parallelism calibration
-0.771481
Date:
4/28/2020
Please move the axes to the
first point.
Cancel
Save and exit
Next
.