GARANT VG Basic pre-setting device
de
ü Adapter/calibration mandrel inserted in the tool mounting.
1. In the measurement software, select or create the adapter/calibration mandrel that was inserted.
Calibrate the adapter
2. Press
en
3. Release the locking of the quick adjustment, move the tower and optical unit into position.
4. Use the infinite fine adjustment to move the camera to the position of the calibration edge of the adapter.
5. View the camera image on the panel PC; the X-value and Z-value must both lie on the calibration edge.
cs
6. Acknowledge the inserted adapter in the measurement software; the measured values will be displayed.
»
The VG Basic is set up and calibration is complete.
»
Measurements can be performed using the calibrated adapter.
da
HSK50
es
HSK63
fi
SK40
SK50
fr
hr
hu
Fig. 6: Start screen of the measurement software, here is an example with adapters already inserted
it
lt
nl
pl
ro
ru
Fig. 7: Acknowledge the inserted adapter
i
SK50 tool mounting without calibration edge. When using with SK50 tools, use a suitable calibration mandrel such as Article
sl
No. 355105 90.