17000 - CO2 Laser Alignment
CO2 Laser Alignment
CO2 laser alignment should be checked
periodically to ensure that the engraver is
operating at optimum performance.
Complete this procedure if any of the
following behaviors are observed:
•
Laser power appears weak in certain
areas of the table, especially when
moving from left to right in the
engraver
•
The engraver is unable to effectively
cut through materials at or around the
recommended settings.
•
Optics in the lens assembly fail
prematurely.
•
If the engraver has been moved to a
new location.
Or if any of these components have been
replaced:
•
Laser tube
•
Mirror #1, #2, or #3
•
X-axis assembly
•
Red dot diode
CO2 Laser Safety
The laser assembly used in the Fusion Pro
CO2 engraver can cause serious and or
permanent injury if safety procedures are
not observed.
Do not fire the CO2 laser with any of the
engraver panels removed.
Do not place any part of your body in the
beam path while the laser is in operation.
Do not defeat the safety interlock system
in the engraver.
It is recommended that eye protection
such prescription eyeglasses, Lexan, or
other polycarbonate safety glasses be
worn by all those present for the duration
of the laser alignment procedure.
2
!
!