Configuration
October 2023
9. Click the Frequency and Digital Output 2 tab. Follow the instructions in the
10. Click the Analog Input 1 (Temperature) button in Inputs tab. Configure the input
11. Click the Analog Input 2 (Pressure) button Inputs tab. Configure the input
12. Click the Digital Input button in Inputs tab. The default Digital Input 1 polarity is set
84
a) Click Apply, after you enter the data to write the parameters to the meter.
to configure the Frequency and Digital Output 2 parameters.
a) Click Apply, after you enter the data to write the parameters to the meter.
parameters including: Source (Live Analog or Fixed), Min and Max input limits
corresponding to 4 mA and 20 mA respectively and the Low and High alarm limits.
a) Click Apply, after you enter the data to write the parameters to the meter.
parameters including: Source (Live Analog or Fixed), Min and Max input limits
corresponding to 4 mA and 20 mA respectively and the Low and High alarm
limits. Select either Gage or Absolute for the type of pressure reading desired. If
a live pressure transmitter is connected, select the type of reading the transmitter
outputs. If Absolute is selected, you must also enter the Atmospheric pressure.
a) Click Apply, after you enter the data to write the parameters to the meter.
to Normal for general purpose or set to Inverted when used for calibration.
a) Click Apply, after you choose the calibration data to write the parameters to
the meter.
Calibration Polarity configuration parameter selections are:
Digital Input 1 Calibrate Active High
Digital Input 1 Calibrate Active Low
Calibration Gating configuration parameter selections are:
Edge gated, active high
Figure 4-9: Gating configuration parameter Edge gated, active high
Edge gated, active low
Figure 4-10: Gating configuration parameter Edge gated, active low
Installation manual
00825-0600-3104
Step 4c
Gas Ultrasonic Flow Meters