Emerson Rosemount 3410 Series Reference Manual

Emerson Rosemount 3410 Series Reference Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Rosemount 3410 Series:

Advertisement

Quick Links

Reference Manual
00825-0200-3104, Rev AB
January 2023
Rosemount
3410 Series Gas Ultrasonic
Flow Meters
Models 3415, 3416 and 3417

Advertisement

Table of Contents
loading

Summary of Contents for Emerson Rosemount 3410 Series

  • Page 1 Reference Manual 00825-0200-3104, Rev AB January 2023 Rosemount ™ 3410 Series Gas Ultrasonic Flow Meters Models 3415, 3416 and 3417...
  • Page 2 Emerson employees. Emerson will not accept your returned equipment if you fail to follow Emerson procedures. Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Emerson Customer Service department.
  • Page 3: Table Of Contents

    Reference Manual Contents 00825-0200-3104 January 2023 Contents Chapter 1  Introduction..........................5 1.1 Typical applications of this product..................5 1.2 Features and benefits of the 3415, 3416 and 3417 models meter........5 1.3 Acronyms, abbreviations and definitions.................6 1.4 MeterLink software........................8 ™ 1.5 Design of Rosemount 3410 series meter................9 1.6 Meter specifications for 3415, 3416 and 3417 models............14 1.7 Preinstallation considerations....................
  • Page 4 Contents Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 5: Chapter 1 Introduction

    Reference Manual Introduction 00825-0200-3104 January 2023 1  Introduction 1.1  Typical applications of this product Rosemount Dual-Configuration 3410 Series Ultrasonic Gas Flow Meters have various ™ configurations that meet a broad range of customer requirements. Each meter comes fully assembled from Rosemount. The technology can be applied to custody transfer, allocation measurement, and check metering applications such as: •...
  • Page 6: Acronyms, Abbreviations And Definitions

    • Ethernet access • On-board LED status indicators • Analog pressure and temperature inputs • Communication via Emerson’s AMS Device Manager and Field Communicator ™ • API Chapter 21 compliant event and data logging (gas meters) • MeterLink (a Windows - based interface software) ™...
  • Page 7 Reference Manual Introduction 00825-0200-3104 January 2023 Acronym or abbreviation Definition C T S Clear-to-Send; the RS-232C hand shaking signal input to a transmitter indicating that it is okay to transmit data - i.e., the corresponding receiver is ready to receive data. Generally, the Request-to-Send (RTS) output from a receiver is input to the Clear-to-Send (CTS) input of a transmitter.
  • Page 8: Meterlink Software

    Watts (power unit) 1.4  MeterLink software MeterLink software has robust features for setting communications parameters, configuring your meter, collecting logs and reports and monitoring the meter health and alarm statuses. MeterLink may be downloaded at no charge from: www.emerson.com/ meterlink. Emerson.com/Rosemount...
  • Page 9: Design Of Rosemount ™ 3410 Series Meter

    3410 series meter ™ Rosemount 3410 Series Gas Ultrasonic Flow Meters are designed to accurately measure products in applications where reliable performance is critical, by measuring the difference in signal transit time with and against the flow across one or more measurement path(s). A signal transmitted in the flow direction travels faster than one transmitted against the flow direction.
  • Page 10 The single check measurement path is positioned at 30 degrees off vertical while the diagnostic path is vertical. All measurement paths measure the difference in signal transit time with and against the flow. Emerson.com/Rosemount...
  • Page 11 Reference Manual Introduction 00825-0200-3104 January 2023 Figure 1-3: Rosemount 3416 and 3415 Gas Ultrasonic Flow Meter design A. Transmitter electronics enclosure (explosion-proof) Optional - Local Display with glass endcap (Figure 1-6) B. Base electronics enclosure (intrinsically safe) C. Meter body with transducer assemblies (T-21, T-22 or T-200) (intrinsically safe) Rosemount 3415 Gas Ultrasonic Flow Meter is a custody meter and check meter in one body.
  • Page 12 Rosemount Model 3415/16/17 Dual Configuration Meters can be connected to customer infrastructure via single Ethernet or Serial connection when Expansion I/O Module is installed and ethernet connection wired locally between Head 1 and Head 2. Expansion IO Module P/N 1-360-03-026 requires Type 4 CPU Module (1-360-03-065) for operation. Emerson.com/Rosemount...
  • Page 13 Reference Manual Introduction 00825-0200-3104 January 2023 Figure 1-5: Dual configuration meter data sharing connections through local network (No Expansion I/O Module) Rosemount Model 3415/16/17 Dual Configuration Meters can be connected to customer infrastructure through separate Ethernet connections to each head. Expansion I/O Module is not required for data sharing.
  • Page 14: Meter Specifications For 3415, 3416 And 3417 Models

    The meter must be fully depressurized and drained before attempting to remove the T-200 transducer assembly. If gas or fluid begins to leak from the T-200 transducer stalk assembly, stop immediately and reinstall T-200 stalk assembly. Failure to comply could cause serious injury or equipment damage. A. Transducer holder Emerson.com/Rosemount...
  • Page 15 Reference Manual Introduction 00825-0200-3104 January 2023  CAUTION ESCAPING GASES OR FLUIDS HAZARD The purchaser of the meter is responsible for the selection of Rosemount components/ ™ seals and materials compatible with the chemical properties of gas flow measurement. Failure to select the suitable meter component/seals may cause escaping gases or liquids, resulting in injury or equipment damage.
  • Page 16 ±0.05% of reading in the specified velocity range from 5% to 100% (Q Velocity range • 100 ft/s (30m/s) with over-range) • 125 fps (38 m/s) on some line sizes • Meter meets or exceeds AGA9 (2007) performance specifications Emerson.com/Rosemount...
  • Page 17 Reference Manual Introduction 00825-0200-3104 January 2023 Table 1-2: Performance specifications Models 3415 and 3416 AGA 9 / ISO 17089 Flow rate values (Metric units) Meter size (DN) 200 to 600 Qmin (m/s) Qt (m/s) 3.048 Qmax (m/s) 30.48 Models 3415 and 3416 AGA 9 / ISO 17089 Flow rate values (US Customary Units) Meter size (in) 8 to 24...
  • Page 18 • ± 1% without a flow calibration (10” and smaller line sizes) • ± 0.7% without a flow calibration (12” and larger line sizes) • ±0.1% with a flow calibration Minimum operating pressure 100 psig (7 bar) Electronic specifications Emerson.com/Rosemount...
  • Page 19 Reference Manual Introduction 00825-0200-3104 January 2023 Table 1-3: Rosemount models 3415, 3416 and 3417 meter specifications (part 2) ™ (continued) Meter Power • 10.4 VDC to 36 VDC • 11 W power consumption (15 W maximum) Serial cable • Belden #9940 or equivalent (22 gauge) —...
  • Page 20 AI-1 and AI-2 are electronically isolated and operate in sink mode. The input contains a series resistance so HART Communicators can ® be connected to configure sensors. A regulated 24 Volt DC power output is available to provide power to the sensors. Emerson.com/Rosemount...
  • Page 21 Reference Manual Introduction 00825-0200-3104 January 2023 Table 1-5: Rosemount models 3415, 3416 and 3417 meter specifications (part 3) ™ (continued) Frequency/Digital Output(s) The outputs have user-configurable selections as either a frequency output or digital status (FODO) (Also see Frequency/Digital outputs). Frequency/Digital Outputs •...
  • Page 22: Preinstallation Considerations

    Electrical safety compliance; UL, CSA, ATEX, IECEx etc. • Civil and structural good practices compliance • Contractual agreements or governmental compliance (or both) • In-situ performance test procedures • Field tested meter health check and flow dynamics diagnostics • Data collection and retention procedures Emerson.com/Rosemount...
  • Page 23: Safety Considerations

    (see 3410 Series engineering drawings). Rosemount 3410 Series GasUltrasonic Meters are INMETRO certified. Refer to the 3410 Series Gas Ultrasonic Flow Meter Tag, INMETRO Certification drawing DMC-006224. Certificate number: UL-BR 16.0144X Marking: Ex db ia IIB T4...T3 Gb...
  • Page 24: Fcc Compliance

    • NEPSI • GOSTR Important Please consult Emerson Flow services for Rosemount products for the complete metrology approvals list. 1.10  FCC compliance This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
  • Page 25 Reference Manual Introduction 00825-0200-3104 January 2023 2. Measurement of Fuel Gas By Turbine Meters, American Gas Association, Transmission Measurement Committee Report No. 7, Second Revision, April 1996 (also referred to as AGA7) 3. Compressibility Factors of Natural Gas and Other Related Hydrocarbon Gases, American Gas Association, Transmission Measurement Committee Report No.
  • Page 26 Introduction Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 27: Chapter 2 Mechanical Installation

    Reference Manual Mechanical installation 00825-0200-3104 January 2023 2  Mechanical installation 2.1  Meter piping, lifting and mounting Refer to the following sections for piping recommendations, lifting with hoist rings and slings, mounting in heated or cooled pipelines and safety warnings and precautions.  CAUTION SURFACE TEMPERATURE HAZARD The meter body and piping may be extremely hot or cold.
  • Page 28 The purchaser of the meter is responsible for the selection of Rosemount components/ ™ seals and materials compatible with the chemical properties of gas flow measurement. Failure to select the suitable meter component/seals may cause escaping gases or liquids, resulting in injury or equipment damage. Emerson.com/Rosemount...
  • Page 29: Meter Components

    Failure to select the suitable meter seals may cause escaping gases or liquids, resulting in injury or equipment damage. Consult your Emerson Flow sales and service representative to ensure you purchase the correct components and seals for your application. 2.2 ...
  • Page 30 During equipment maintenance, disconnect power before opening transmitter or base electronics. Clean cover joints before replacing. • DO NOT substitute meter components. Component substituting may compromise the intrinsic safety. Failure to comply could result in severe injury to personnel or cause damage to the equipment. Emerson.com/Rosemount...
  • Page 31 Reference Manual Mechanical installation 00825-0200-3104 January 2023 The 3417 dual 4-path ultrasonic meter components are shown below. Figure 2-1: Rosemount 3417 Flow Meter assembly A. Explosion-proof transmitter enclosure (CPU Module, Power Supply, I.S. Barrier Board Backplane board) (Optional: glass endcap for Local Display) B.
  • Page 32 Backplane board) (Optional: glass endcap for Local Display) B. Intrinsically-safe base enclosure includes Acquisition Module C. Meter - body and transducer assemblies and cables D. Flange stabilizers (2) Model 3415 does not have the vertical bounce path (see Rosemount 3415 description) Emerson.com/Rosemount...
  • Page 33 Reference Manual Mechanical installation 00825-0200-3104 January 2023 Figure 2-3: Transmitter electronics enclosure with optional local display and glass endcap A. Transmitter electronics enclosure with glass endcap B. Local display Rosemount 3415, 3416 and 3417 GUSM...
  • Page 34: Piping Recommendations

     CAUTION SUNSHIELD PROTECTION Install a sunshield to prevent prolonged exposure to direct sunlight in extreme climates. Failure to shield the meter may result in exceeding the process temperature range and damage transmitter electronics. Emerson.com/Rosemount...
  • Page 35 Reference Manual Mechanical installation 00825-0200-3104 January 2023 NOTICE For optimal flow measurement conditions, Rosemount suggests the piping configurations ™ below. Regardless of the configuration selected, the user agrees to accept full responsibility for the site piping design and installation. Flow conditioning is recommended for best measurement results •...
  • Page 36 Gas Ultrasonic Flow Meter link, click the Documentation tab, expand the Data Sheets - Bulletins - Catalogs tab, then select the Data Sheet. Rosemount 3410 Series Ultrasonic Gas Flow Meters should be mounted in horizontal piping with the chord paths horizontal.
  • Page 37: Pre-Installation Inspection

    1. Ensure flange sealing faces are undamaged. 2. Movement of components that should be rigid. If any damage is found, contact Emerson Flow services before putting meter into service. Refer to the Emerson Flow services contact information on the back cover of this manual. 2.4.1 ...
  • Page 38 2-12) in the Rosemount Ultrasonic Meter flange tapped holes to aid in lifting or maneuvering the unit. Operators SHALL NOT use other Hoist Rings that do not fully seat flush with the counter bore on the top of the meter flanges. Emerson.com/Rosemount...
  • Page 39 Reference Manual Mechanical installation 00825-0200-3104 January 2023 Figure 2-10: Meter end flange with tapped flat-counterbore hole for hoist ring A. Plug bolt B. Flat Counterbore surface Figure 2-11: Safety approved hoist ring Rosemount 3415, 3416 and 3417 GUSM...
  • Page 40 The slings must be of equal length. Each sling must have a load rating that equals or exceeds the hoist ring load rating. The angle between the two slings going to the hoist rings must never exceed 90 degrees or the load rating of the hoist rings will be exceeded. Emerson.com/Rosemount...
  • Page 41 Reference Manual Mechanical installation 00825-0200-3104 January 2023 Figure 2-13: Correct sling attachment for Shell Shroud meters 9. NEVER allow the slings to contact the electronics enclosure. Damage to the enclosure may occur. Use a spreader bar with the slings to prevent contact with the electronics enclosure and the base enclosure (see Safety precautions using appropriate rated lifting...
  • Page 42 Obtain safety engineered swivel hoist rings A list of approved manufacturers of safety engineered hoist rings is below: • American Drill Bushing Company (http://www.americandrillbushing.com) • Carr Lane Manufacturing Company (http://www.carrlane.com) Emerson.com/Rosemount...
  • Page 43 The following instructions are intended to provide general guidelines for using proper lifting slings when lifting a Rosemount 3410 Series Gas Ultrasonic Flow Meter by itself. These instructions are intended to be followed in addition to your company's standards or the DOE-STD-1090-2004 Hoisting and Rigging standard if such company standards do not exist.
  • Page 44 J3 on the Acquisition Module, return the electronics enclosure to its original position, replace the bolts, and secure the enclosure in place. Lifting the meter with the upper enclosure installed but without the bolts installed, may cause the electronics to fall and cause personal injury or electronics damage. Emerson.com/Rosemount...
  • Page 45: Mounting Requirements In Heated Or Cooled Pipelines

    Reference Manual Mechanical installation 00825-0200-3104 January 2023 Figure 2-16: Incorrect sling attachment 8. NEVER apply shock loads to the meter. Always lift the meter gradually. If shock loading ever occurs, the slings must be inspected per manufacturer's procedures prior to being placed in any further service. 2.5 ...
  • Page 46 Mechanical installation Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 47: Chapter 3 Electrical Installation

    Reference Manual Electrical installation 00825-0200-3104 January 2023 3  Electrical installation 3.1  Cable length TTL mode The maximum cable length is 2000 feet when the Digital Output “TTL” mode is selected. 3.2  Cable length Open Collector mode For the Digital Output “open collector” mode, the maximum cable length depends on the cable parameters, pull-up resistance used, the maximum frequency to output, and frequency input parameters being driven.
  • Page 48 DO NOT connect digital grounds to the ground lugs. Figure 3-1: Internal Transmitter Electronics Enclosure chassis ground A. Transmitter Electronics Enclosure ground lug Emerson.com/Rosemount...
  • Page 49: Conduit Seals

    Reference Manual Electrical installation 00825-0200-3104 January 2023 Figure 3-2: External ground lug A. External ground lug 3.4  Conduit seals Conduit seals are required for meter installations in hazardous environments. Adhere to safety instructions to protect personnel and equipment.  WARNING EXPLOSION HAZARD To reduce the risk of an explosion or fire, conduit runs must have a sealing fitting connected within 457.2 mm (18 inches) of the enclosure.
  • Page 50 3. Remove the end cap nearest the conduit entry to gain access to the transmitter electronics. 4. Pull the wires into the electronics enclosure. Complete the field connection wiring as shown in Figure 3-3. 5. Complete the field connection wiring and apply electrical power to the system. Emerson.com/Rosemount...
  • Page 51 Reference Manual Electrical installation 00825-0200-3104 January 2023 Figure 3-3: Electronics field wiring - upper terminal block, switches, ground lug - Type 2 CPU Module A. Conduit wiring entry (four entries) c. Analog In B. Switches: • Analog In (AI1) 1. Port A —...
  • Page 52 C. Power in (10.4 VDC - 36 VDC) • AO1- D. Fuse cover c. Serial COMs (RS-323, RS-485) • RS-232: RTS, TX, RX, CTS • RS-485: TX+, TX-, RX+, RX- (4-wire Full Duplex) • RS-485: TX+, TX- (2-wire Half Duplex) Emerson.com/Rosemount...
  • Page 53 Reference Manual Electrical installation 00825-0200-3104 January 2023 Figure 3-5: Electronics field wiring - upper terminal block, switches, ground lug - Type 4CPU Module A. Conduit wiring entry (four entries) b. Analog In B. Switches: • Analog In (AI1) 1. Port A —...
  • Page 54 6. Set or configure the meter operating parameters using MeterLink. For additional installation information refer to the system wiring diagram (see ), MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start Manual (00809-0100-7630) and use the MeterLink Field Setup Wizard to complete the configuration. Emerson.com/Rosemount...
  • Page 55: Security Seal Installation

    Reference Manual Electrical installation 00825-0200-3104 January 2023 7. Verify the field connections are working correctly. Allow the system to run for the time specified by the customer (usually one week) and an electrician has fully tested the connections. After the Acceptance Test is witnessed and approved, seal the conduit.
  • Page 56 2. Install the security seal wire into and through one of the two holes in the end cap. a) Choose holes that minimize counterclockwise rotation of the end cap when the security wire is taut (maximum wire diameter 0.078-in.; 2.0 mm). Emerson.com/Rosemount...
  • Page 57 Reference Manual Electrical installation 00825-0200-3104 January 2023 Figure 3-8: Transmitter electronics enclosure security seals A. Transmitter electronics enclosure endcap B. Security wire seals 3. Adjust the security wire, removing all slack and thread into the lead seal. 4. Crimp lead seal and cut wire ends to remove excess wire. 3.5.2 ...
  • Page 58 The transducer assemblies can be protected from tampering by securing the shrouds over the transducers with wire seals as follows. Procedure 1. Locate the shroud pin on the meter body. This pin will hold the appropriate shroud in place while the mating side is brought into place. Emerson.com/Rosemount...
  • Page 59 Reference Manual Electrical installation 00825-0200-3104 January 2023 Figure 3-10: Latch pin and Shroud recesses A. Shroud pin on the meter body 2. Hook the appropriate shroud over the pin, ensuring the transducer cabling is within the shroud. Care needs to be taken not to pinch the cables between the shroud recesses and shroud as the shroud is fitted into place.
  • Page 60: Sealing The Unit

    (approximately one to two weeks) before the unit is sealed. This allows time to verify all electrical connections are correct, that the meter is accurately measuring flow and that the meter meets the customer’s installation requirements. See Startup for systems that use explosion-proof conduit Startup for systems that use flame-proof cable. Emerson.com/Rosemount...
  • Page 61: Chapter 4 Configuration

    Reference Manual Configuration 00825-0200-3104 January 2023 4  Configuration After the mechanical and electrical installation is complete use the following to install MeterLink in order to establish connection with the meter to perform final configuration ™ and verify meter performance. 4.1  Set up the MeterLink ™...
  • Page 62 192.168.135.100 for Head 1 and 192.168.135.101 for Head 2. No changes to configuration in either meter is required to setup connection. Figure 4-1: Meter Directory settings example for Ethernet Connection with DHCP disabled Emerson.com/Rosemount...
  • Page 63: Field Setup Wizard

    Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-2: Meter Directory settings example for Ethernet Connection with DHCP Enabled 4.2  Field Setup Wizard Procedure 1. Use the Field Setup Wizard-Startup in MeterLink and select the checkboxes ™ that allow proper configuration for your meter (Temperature, Pressure, Meter Corrections, Meter Outputs, Gas chromatograph setup, Continuous flow analysis and View local display setup).
  • Page 64 ™ chromatograph using Modbus TCP/IP. Choose Port as Ethernet. • GC protocol: select the protocol for which the GC is configured. The Rosemount ™ Gas Ultrasonic meter uses 7 data bits, Even parity, and 1 stop bit for ASCII Emerson.com/Rosemount...
  • Page 65 Reference Manual Configuration 00825-0200-3104 January 2023 Modbus and 8 data bits, No parity, and 1 stop bit for RTU Modbus. This option will be enabled only when a serial port is selected. • GC baud rate: select the baud rate for which the GC is configured. This option will be enabled only when a serial port is selected.
  • Page 66 Local display labels, descriptions and valid units QFLOW — Uncorrected volume flow rate • ACF – Actual Cubic Feet • ACM – Actual Cubic Meters • MACF – Thousand Actual Cubic Feet • MACM –Thousand Actual Cubic Meters TDYVL — Current day’s forward uncorrected volume Emerson.com/Rosemount...
  • Page 67 Reference Manual Configuration 00825-0200-3104 January 2023 (continued) Table 4-1: Local display labels, descriptions and valid units Local display labels, descriptions and valid units • +ACF – Actual Cubic Feet • +ACM – Actual Cubic Meters • +MACF – Thousand Actual Cubic Feet •...
  • Page 68 -MSCM – Thousand Standard Cubic Meters VEL — Average flow velocity • Ft/S – Feet per Second • M/S – Meters per Second SOS — Average sound velocity • Ft/S – Feet per Second • M/S – Meters per Second Emerson.com/Rosemount...
  • Page 69 Reference Manual Configuration 00825-0200-3104 January 2023 (continued) Table 4-1: Local display labels, descriptions and valid units Local display labels, descriptions and valid units TEMP — Flow-condition temperature • DEGF – Degrees Fahrenheit • DEGC – Degrees Celsius PRESS — Flow-condition pressure •...
  • Page 70: Using Ams Device Manager To Configure The Meter

    3410 Series Gas Ultrasonic Flow Meter. 2. Use the Filter Results By categories to narrow-down your search. a) Select the check box for HART under Communication Protocol. b) Search and select the option Emerson Rosemount Industries under the ™ Brand/Manufacturer category.
  • Page 71 Reference Manual Configuration 00825-0200-3104 January 2023 f) Search and select the desired AMS revision under Host System Revision Figure 4-3: AMS Device search result 3. Click the hyperlink. The file download dialog displays. Click the Save button to save the files to your host system. You may use the default download location or change the directory.
  • Page 72 7. Start the AMS Device Manager using a laptop or PC. 8. Enter login credentials and click OK to launch the application. 9. Click the Configure tab, and then select Guided Setup, Manual Setup or Alert Setup. Figure 4-6: AMS Device Manager Emerson.com/Rosemount...
  • Page 73 Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-7: AMS Device Manager - Overview 4.3.2  AMS Device Manager - Guided setup The Guided setup wizard provides configuration parameter settings for the meter. The Guided Setup is a subset of the Manual Setup parameters. Figure 4-8: AMS Device Manager - Guided Setup Note Before writing configuration changes to your meter, ensure you have saved the...
  • Page 74 Reverse Flow. Click the OK button to return to the Overview page. 7. Click the Service Tools tab on the Overview page and select the Variables tab. The Flow Data, Path Information, Flow Totals, and All Variables data is populated after you are connected to the meter. Emerson.com/Rosemount...
  • Page 75 Reference Manual Configuration 00825-0200-3104 January 2023 a) Click the Flow Data tab and view the Flow Direction (Forward or Reverse), Average Flow and Average Sound Velocities values. b) Click the Path Information tab and view the Chord performance, Gain, SNR (Signal to Noise Ratio) Signal strength (mV), and Noise (mV). c) Click the Flow Totals tab to view the volume totals (forward and reverse uncorrected volume).
  • Page 76 AMS Device Manager - Manual setup Use the Manual Setup wizard to configure the meter’s parameters. See Figure 4-6 Figure 4-7 from the AMS Device Manager Configure menu click Manual Setup. Figure 4-11: AMS Device Manager - Configure Manual Setup Emerson.com/Rosemount...
  • Page 77 Reference Manual Configuration 00825-0200-3104 January 2023 Procedure 1. Remove security wires from the endcap and the Bracket/Cover hex head bolts that secures the Base Enclosure if they are installed. 2. Disable the Write Protect switch in the CPU Module to write any of the following configuration parameters to your meter.
  • Page 78 Figure 4-12: Gating configuration parameter Edge gated, active high — Edge gated, active low Figure 4-13: Gating configuration parameter Edge gated, active low — State gated, active high Figure 4-14: Gating configuration parameter State gated, active high — State gated, active low Emerson.com/Rosemount...
  • Page 79 Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-15: Gating configuration parameter State gated, active low 13. Click the Alert Setup tab (from the main Configuration page). Figure 4-16: Configure Flow Analysis Alert 14. Click the Flow Analysis tab to select Configure Reverse Flow Detection, if desired. The default setting is Disabled.
  • Page 80 Figure 4-17: AMS Device Manager - Service Tools Alerts b) If you change the device configuration, a confirmation dialog displays and prompts you to write the changes to the meter. Click Yes to write the changes to the meter or click No to cancel pending changes. Emerson.com/Rosemount...
  • Page 81 Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-18: Configuration changes dialog c) Click the Service Tools → Variables tab. The Variables page displays tabs for the device’s Flow Data, Path Information, Flow Totals, and All Variables. Figure 4-19: AMS Device Manager - Service Tools The Service Tools → Flow Data page includes charts for flow and sound velocities.
  • Page 82 Maintenance - Fix Soon and Advisory according to the severity level; 1-6. Severity 1 is the highest and 6 is the lowest level. g) Click the Service Tools → Trends tab to display the device variables (uncorrected volume flow rate, pressure and temperature) trends. Emerson.com/Rosemount...
  • Page 83 Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-21: AMS Device Manager - Service Tools Trends Primary and Secondary variables display real-time uncorrected volume flow rate trends. The third and fourth variables charts displays trends for temperature and pressure. 16. Click the Service Tools → Routine Maintenance tab. Click Analog Output 1 Trim to perform a digital to analog trim adjustment of the first milliampere output.
  • Page 84: Using A Field Communicator To Configure The Meter

    Using a Field Communicator to configure the meter Prerequisites • Emerson Field Communicator software, license, installation guide and user manual available on the Emerson Asset Optimization Field Communicator website: www.emerson.com/en-us/catalog/ams-475-field-communicator • Rosemount HART Device Description (HART DD) installed for the meter ™...
  • Page 85 Reference Manual Configuration 00825-0200-3104 January 2023 Figure 4-22: 3416 transmitter field wiring conduit entries 5. Wire Analog Input 1 (AI1) and Analog Output 1 (AO1) as shown in Figure 4-23 Engineering drawings, drawing DMC-005324. Figure 4-23: Field Communicator wiring diagram for the 3410 Series electronics 6.
  • Page 86: Security Seals For The Meter

    Meter → Manage Users dialog box in MeterLink. If the Rosemount 3410 Series Electronics will be connected to a network, please read the security recommendations found in Cybersecurity and network communications in the Rosemount 3410 Series Gas Ultrasonic Flow Meters: Operations Manual (00809-0300-3104).
  • Page 87: Chapter 5 Engineering Drawings

    Reference Manual Engineering drawings 00825-0200-3104 January 2023 5  Engineering drawings 5.1  3410 Series engineering drawings This appendix contains the following engineering drawing(s) for the ultrasonic meter: DMC-005324 Rosemount 3410 Series Ultrasonic Gas Flow Meter ™ System Wiring Diagram Rosemount 3415, 3416 and 3417 GUSM...
  • Page 88 Engineering drawings Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 89 Reference Manual Engineering drawings 00825-0200-3104 January 2023 Rosemount 3415, 3416 and 3417 GUSM...
  • Page 90 Engineering drawings Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 91: Open Source Licenses

    6.1  List of source codes for executable files For a copy of the source code covered under the open source licenses indicated in this appendix, please contact flow.support@emerson.com. 6.1.1  GNU General Public License For more details about GNU GPL (General Public License), follow the link below: www.gnu.org...
  • Page 92 License. (Exception: if the Program itself is interactive but does not normally print such an announcement, your work based on the Program is not required to print an announcement.) Emerson.com/Rosemount...
  • Page 93 Reference Manual Open source licenses 00825-0200-3104 January 2023 These requirements apply to the modified work as a whole. If identifiable sections of that work are not derived from the Program, and can be reasonably considered independent and separate works in themselves, then this License, and its terms, do not apply to those sections when you distribute them as separate works.
  • Page 94 FOR THE PROGRAM, TO THE EXTENT PERMITTED BY APPLICABLE LAW. EXCEPT WHEN OTHERWISE STATED IN WRITING THE COPYRIGHT HOLDERS AND/OR OTHER PARTIES PROVIDE THE PROGRAM "AS IS" WITHOUT WARRANTY OF ANY KIND, EITHER EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF Emerson.com/Rosemount...
  • Page 95 Reference Manual Open source licenses 00825-0200-3104 January 2023 MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE PROGRAM IS WITH YOU. SHOULD THE PROGRAM PROVE DEFECTIVE, YOU ASSUME THE COST OF ALL NECESSARY SERVICING, REPAIR OR CORRECTION.
  • Page 96 Application does not supply the function or data, the facility still operates, and performs whatever part of its purpose remains meaningful, or "b) under the GNU GPL, with none of the additional permissions of this License applicable to that copy. Emerson.com/Rosemount...
  • Page 97 Reference Manual Open source licenses 00825-0200-3104 January 2023 3. Object Code Incorporating Material from Library Header Files. The object code form of an Application may incorporate material from a header file that is part of the Library. You may convey such object code under terms of your choice, provided that, if the incorporated material is not limited to numerical parameters, data structure layouts and accessors, or small macros, inline functions and templates (ten or fewer lines in length), you do both of the following:...
  • Page 98 ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE. 6.1.4  M.I.T License For more details about the Open Source MIT license or the Open Source Initiative follow ™ the link below: www.opensource.org/licenses/mit-license.php Emerson.com/Rosemount...
  • Page 99 Reference Manual Open source licenses 00825-0200-3104 January 2023 The MIT License Copyright (c) <<year>> <<copyright holders>> Permission is hereby granted, free of charge, to any person obtaining a copy of this software and associated documentation files (the "Software"), to deal in the Software without restriction, including without limitation the rights to use, copy, modify, merge, publish, distribute, sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is furnished to do so, subject to the following conditions:...
  • Page 100 Open source licenses Reference Manual January 2023 00825-0200-3104 Emerson.com/Rosemount...
  • Page 101 Reference Manual 00825-0200-3104 January 2023 Rosemount 3415, 3416 and 3417 GUSM...
  • Page 102 Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.

This manual is also suitable for:

Rosemount 3415Rosemount 3416Rosemount 3417

Table of Contents