Figure 2-44 Through–Bolt Supplied (Slab Type Floor Construction)
Floor requirements when using provided floor anchors
The maximum pullout force per provided anchor was calculated assuming:
• A concrete compression strength of 17.24 MPa at 28 days (which is the minimum required
compression strength).
• Anchors installed to the required hole depth of 165.1 mm (6.5 in) minimum, and
• Center of anchor hole to concrete edge distance 79.4 mm (3.12 in).
Make sure to obtain data on compression strength of the concrete before using floor anchors.
Pan Type Floor Construction Requirement
For Pan type floor construction, steel channels must be designed by a local structural engineer to span
floor joists. See
Figure 2-45 Through–Bolt Supplied (Pan Type Floor Construction) on page
NOTE
For specific floor preparation procedures, refer to Single Plane and Biplane Innova Systems
Pre–Installation Kit Installation Procedures.
Figure 2-45 Through–Bolt Supplied (Pan Type Floor Construction)
5750182-1EN 3
TM
Innova
IGS 6
Equipment Requirements