Special Construction Requirements
Table 3-1
Emissions
Radio–Frequency
emissions
CISPR11
Harmonic emis-
sions
IEC 61000-3-2
Voltage fluctua-
tions / flicker emis-
sions
IEC 61000-3-3
NOTE
(1)
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or
re-orienting the equipment.
Electromagnetic Immunity
Electromagnetic Immunity IEC 60601-1-2
The Innova system is intended for use in the electromagnetic environment specified below.
The purchaser or user of the Innova system should assure that it is used in such an environment.
120
Test Compliance
The system uses Radio Frequency energy only for its internal function.
Therefore, its Radio Frequency emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Group1 Class A
(1)
limits
The system is suitable for use in all establishments other than domes-
tic and those directly connected to the public low voltage power sup-
ply network that supplies buildings used for domestic purposes.
Not applicable
The system is suitable for use in all establishments other than domes-
tic and those directly connected to the public low voltage power sup-
ply network that supplies buildings used for domestic purposes.
Not applicable
The system is suitable for use in all establishments other than domes-
tic and those directly connected to the public low voltage power sup-
ply network that supplies buildings used for domestic purposes.
TM
Innova
IGS 6
Electromagnetic Environment
5750182-1EN 3