GE Innova IGS 6 Preinstallation Manual page 94

Table of Contents

Advertisement

Layout Constraints
Table 2-11 System Layout constraints Technical and Exam room on same floor
Location
Technical room
Technical room
Technical room
Technical room
Technical to Exam
room
Technical to Exam
room
Technical to Exam
room
Technical to Exam
room
Technical to Exam
room
5750182-1EN 3
Constraint Name
Cabinets location
Tube chillers location
Detector conditioner
location
PDB location
Distance between
Frontal Gantry foot
and C1 or C2 entry.
Distance between
Frontal Gantry foot
and tube chiller
Distance between
Frontal Gantry foot
and Detector Condi-
tioner
Distance between
Lateral Gantry wall-
box and C1 or C2 en-
try
Distance between
Lateral Gantry wall-
box and tube chiller
TM
Innova
IGS 6
Parameter
Specification
Maximum rela-
9 m (29 ft, 6 in)
tive distance
The Tube Chiller
shall be no more
than 5 m (16
feet) above or 8
m (26 feet) be-
low the upper
position of the
X-Ray Tube
In tech room
Max. 3 m (9 ft,
10 in) below Lat-
eral Gantry rails
In Tech Room
Relative distance 19.5 m (64 ft)
Relative distance 22 m (72 ft)
Relative distance 22 m (72 ft)
Max relative dis-
10.5 m (34 ft, 5
tance (proj. on
in)
floor)
Max relative dis-
11 m (36 ft, 1 in)
tance (proj. on
floor)
Equipment Requirements
Comment
Make sure to provide
enough space for storing ex-
tra cable length (e.g. Plinth)
Tech and Exam room on
same floor. 22 m (72 ft) for
HV cables.
Tech and Exam room on
same floor
Tech and Exam room on
same floor
Tech and Exam room on
same floor
Tech and Exam room on
same floor
83